SG-1000

Specific Gravity Tester

User Manual

The 

BEST

 Battery Testing Equipment. The 

BEST

 Customer Support. 

PERIOD. 

www.eepowersolutions.com

Copyright © 2012 Eagle Eye Power Solutions, LLC

V1.0

Summary of Contents for SG-1000

Page 1: ...SG 1000 Specific Gravity Tester User Manual The BEST Battery Testing Equipment The BEST Customer Support PERIOD www eepowersolutions com Copyright 2012 Eagle Eye Power Solutions LLC V1 0...

Page 2: ......

Page 3: ...shed onto the skin or clothing rinse with sufficient amounts of cold water If it gets into an eye flush with cold clean water for about 15 minutes and seek immediate medical attention Remove the batte...

Page 4: ...mes and Functions of Parts 4 2 Inserting the Nozzle 6 3 Inserting the Battery 7 4 Attaching the Wrist Strap 8 5 Zero Setting 9 6 Measurement of Sample 11 7 Explanation of Error Messages 13 8 Temperatu...

Page 5: ...Eagle Eye SG 1000C User Manual 4 1 Names and Functions of Parts...

Page 6: ...setting 4 START OFF key Press to take measurement To turn off the instrument press and hold down the key for 2 seconds 5 Battery Cover Remove to replace battery 6 Nozzle Cap Used to protect tip of noz...

Page 7: ...e is packed separately from the Main unit Insert the Nozzle to Main Unit following the procedure described below 1 Push the Nozzle into the hole located 2 Then fit the nozzle into the slit by at the b...

Page 8: ...itself is not guaranteed and be sure to use a new battery if its voltage is low 3 Close the Battery Cover and using a Screwdriver turn the locking pin to the CLOSE position Note The SG 1000 does not h...

Page 9: ...SG 1000C User Manual 8 4 Attaching the Wrist Strap Attach the wrist strap to the hook located on the lower side of the Main Unit When measuring samples loop the strap around your wrist to avoid any a...

Page 10: ...at the Sample Chamber has been filled with water If bubbles are present in the Sample Chamber operate the Tube Pump again to refill water into the Sample Chamber 2 Dip the tip of the Nozzle in water a...

Page 11: ...Chamber Allow the unit to equilibrate to the ambient temperature before taking measurements Zero setting can be performed anytime Using water the unit will be calibrated to zero Use distilled water fo...

Page 12: ...d squeeze and release the Tube Pump a few times to draw the sample into the sample chamber Once the sample is drawn into the unit Drain it completely and then refill the sample into the unit again Thi...

Page 13: ...er for a few days the prism surface will be soiled and consequently cause errors during measurement To clean the prism surface use a soft tissue soaked with ethyl or isopropyl alcohol If the electroly...

Page 14: ...when performing zero set LLL Sampling error Electrolyte in the Sample Chamber is insufficient to cover the prism surface or air bubbles may be present HHH Range over error The sample measured is out o...

Page 15: ...ce 1 To replace the Tube Pump slide it to the left to release and replace it with a new one 2 After measurement fill and drain the Sample Chamber with distilled or tap water several times to clean Com...

Page 16: ...The Sample Chamber Sample Hook and or Tube Pump may have been freed from the normal position Make sure the Sample Chamber Hook is properly attached to the unit Adjust the Sample Chamber and Tube Pump...

Page 17: ...Eagle Eye SG 1000C User Manual 16...

Page 18: ...the ambient temperature and the sample temperature close to equilibrium Does the battery need replacement Is the O ring damaged in the Sample Chamber Is the Tube Pump damaged Is the Nozzle damaged Is...

Page 19: ...ttery Refer to page 8 on how to replace the battery and other remarks The unit uses a 9V DC battery 006P Is the O ring damaged in the Sample Chamber Remove the Sample Chamber cover from the unit see p...

Page 20: ...the Nozzle is securely attached Leakage of liquid or air can be observed if the Tube Pump Nozzle or cover has been damaged or the Nozzle is loose If damage is found replace the damaged part Two spare...

Page 21: ...he sample before pressing the START key 11 Specifications Model SG 1000 Detection method Optical refraction Measurement range Specific gravity of electrolyte 1 000 to 1 300 Resolution 0 001 Measuremen...