The material contained in this document is the property of Electronics & Innovation Ltd., it is
subject to change without notice.
March 2007
10
Revision B
Performance limits quoted are for guidance only and should not be taken for
guaranteed performance specifications unless they are also quoted in the
Specification Section 1.2.
4.2 PERFORMANCE CHECKS
To determine the amplifier’s performance carry out the following procedure.
4.2.1 Initial Check
The following check can be made after repair and adjustments or
whenever the condition of the unit is in question.
a.
Connect AC power supply. Switch on power and observe that the
display initializes.
b.
Connect a sweep generator (HP 8601 or similar) capable of sweeping
the frequency range 10 KHz to 5 MHz, to the input connector.
c.
Adjust the output level of the sweep generator so that a 50 ohm video
detector connected at the output of the unit will not be damaged by
excessive power output. (Reference section 4.4.1 for set up.)
d.
Observe the gain versus frequency ripple on an oscilloscope calibrated
in decibels. The gain variation must be not more than +/- 1.5 dB over
the frequency range.
e.
Connect a calorimetric power meter (HP435B or equivalent) through a
30 dB 500 watt attenuator to the output connector. Adjust the input CW
signal to any frequency between 30 kHz and 5 MHz for 400 watts
output.
f.
Observe the harmonic distortion of the output, properly attenuated, on
a spectrum analyzer. The harmonic components contributed by the
amplifier should be better than 10 dB down from the fundamental.
(a)
If the above items are found to be outside of the specification, check
the spectral content of the input signal. If this is a pure signal then
the unit needs to be returned to the factory for service.