Fig 4-16
4.2.17 Correction Function
Correction
Function
Description
Typical Use
Model
Open
circuit
correction
Compensation
for
stray
admittance
caused by test fixture
High impedance
measurement
Short
circuit
correction
Compensate for the
residual impedance
caused by the test
fixture
Low impedance
measurement
Open/sho
rt circuit
correction
Compensate for the
stray admittance and
residual impedance
caused by the test
fixture
Precision measurement
----
Step 1: Press the
key to enter the correction interface. The interface is as shown in
Fig.4-17. If you do not want to perform correction, press the Main parameter function
key
to exit. In this mode, only the
key and the Main parameter function key
are valid. (Open circuit: OPEN, short circuit, SHORT, neither: ERROR). Connect the test
fixture to the gold plated short circuit board (short circuit) or disconnect the outside of the fixture
(open circuit).