NOTE:
Users can also calibrate either point.
To calibrate 33% only, press key to
exit when 33% calibration is completed.
To calibrate 75% only, press or
key within 5 minutes while initializing
33% calibration. And it skips 33% and
enters 75% calibration mode.
" CAL" icon and calibrating value (75.2%
if at 25°C) are blinking on the LCD with
current temp. at the left. Meter is now
calibrating. Wait about 60 minutes until
blinking stops, then calibration is
completed and it returns to normal mode.
? Error messages
E01: CO sensor damaged.
2
E02: The value is under range.
E03: The value is over range.
E04: The original data error results in
this error (RH, DP, WB)
E07: Too low voltage to measure CO
2
.
Check if the adaptor output is 12V.
E11: Retry humidity calibration.
E17: Retry CO calibration.
2
E31: Temperature sensor damaged.
E34: Humidity sensor damaged.
? Can’t power on
Check whether the adaptor is well
plugged.
TROUBLE SHOOTING
? Slow response
Check whether the air flow channels on
the rear were blocked.