Message

 

What it means? 

What to do?

 
 
 

 

 

 

The “Ht” and thermometer 
icon appears. Temperature is 
too high, outside the required 
range of 10°C - 40°C (50°F - 
104°F). This alerts users that 
an incorrect result may occur 
if the test continues. 

Relocate the meter to a location 
with temperature between 
10°C - 40°C (50°F - 104°F).

 

 

The “Lt” and thermometer icon 
appears. Temperature is too 
low, outside the required range 
of 10°C - 40°C (50°F - 104°F). 
This alerts users that an 
incorrect result may occur if 
the test continues. 

Relocate the meter to a location 
with temperature between   
10°C - 40°C (50°F - 104°F).

 

 

 

 

Summary of Contents for Voice

Page 1: ...Voice Self Monitoring Blood Glucose System User s Manual Please read this User s Manual thoroughly before using your blood glucose meter...

Page 2: ...r choosing the EASYMAX Voice SMBG System Intended Use The EASYMAX Voice Self Monitoring Blood Glucose Test System is intended for the quantitative measurement of glucose in fresh capillary whole blood...

Page 3: ...ccessories Blood Glucose Meter Alkaline Battery 2 ct Lancets 10 pcs Lancing Device User s Manual Warranty Card Carrying Case AST Lancing Device Cap Blood Glucose Test Strips 25 pcs Level 1 Control Sol...

Page 4: ...r your comfort Do you need help If you have questions or need assistance please contact your healthcare professional or visit EASYMAX Website www easymaxdiabetescare com for diabetes management tools...

Page 5: ...dings Disassembling the meter will also void the warranty Always keep the meter clean and store it in a safe place Protect the meter from direct sunlight to ensure a longer lifespan You should not sto...

Page 6: ...products or objects that come in contact with human blood even after cleaning should be handled as if capable of transmitting a viral disease Consult with your healthcare professional before testing...

Page 7: ...t away If you have followed the steps in the user s manual but still have symptoms that do not seem to match your test results or if you have questions please contact your healthcare professional Plea...

Page 8: ...Use by Temperature limitation Manufacturer Catalogue number Serial number Control Sufficient for Authorized representative in the European Community Blood glucose test result in mg dL Keep away from s...

Page 9: ...er Meal Result Mute Record Average display Volume Control Solution Insert strip English Temperature Espa ol Error AM Memory PM Apply Blood 12hr Unit Battery When shows up in any section of user s manu...

Page 10: ...e Please set the date Setting Time Please set the time Meter is ready to insert test strips Please insert a test strip When the meter is ready to apply blood Please apply blood into the strip Blood Gl...

Page 11: ...Please test again and if you get the same result please contact your healthcare professional immediately Press C button to delete a memory Please confirm the deletion Memory was deleted Memory deleted...

Page 12: ...on Testing 20 Why Run A Control Solution Test 20 About The Control Solutions 21 Running A Control Solution Test 22 Understanding Control Solution Test Results 24 Chapter 3 Testing Your Blood Glucose 2...

Page 13: ...emory Storing Test Results 39 Viewing And Deleting Test Results 40 Meter Setup Using The Set Mode 41 Chapter 5 Maintenance And Troubleshooting 42 Inserting Batteries 42 Cleaning Your Meter 43 Cleaning...

Page 14: ...0 memory adjust setting and scroll through results Test Strip Slot Insert test strip here Power Button Press to confirm the setting Battery Door Flip open the battery door by pushing the tab in the di...

Page 15: ...on Area Electrode Lancing Device Trigger button Press the trigger button to activate the lancing device Carrier Insert the lancet into the carrier Adjustable tip Select the desired penetration depth H...

Page 16: ...pen the battery door on the back of the meter by pushing the tab in the direction of the arrow and pulling the door up 2 Insert two batteries You will hear greeting music 3 Put the battery door back i...

Page 17: ...se meter is set at mg dL 3 Press Power Button to turn on the meter and then press C button to set the language volume time and date 1 Insert new batteries the meter turns on Press M button to select E...

Page 18: ...ate in your meter is important if you intend to use the meter memory 1 Press M button and C button to set the year Press power button to confirm the setting 2 Repeat step 1 to set the date 3 Repeat st...

Page 19: ...to check the expiration date on the test strip bottle The test strip will be expired either in six months after the date of the bottle is opened or till the expiration date printed on the bottle Store...

Page 20: ...r meter and test strips are working properly to give reliable results You should run the control solution tests when You think the meter or test strips may be working incorrectly You drop the meter Yo...

Page 21: ...chever comes first Do not use a control solution that is past the expiration date Control solutions can stain clothing If you spill it wash your clothes with soap and water Close the bottle tightly af...

Page 22: ...of the bottle with a tissue 5 Squeeze the bottle until a tiny drop forms at the tip of the bottle and place solution on top of bottle cap Setting the meter in control solution mode will prevent the c...

Page 23: ...ing falls within the range printed on the test strip bottle 9 Press power button to repeat the test result Or you can also press C button or M button to set the volume before repeating the result 10 R...

Page 24: ...n the control solution result is inside the range on the test strip bottle your test strips and your meter are working properly If your control solution result is not inside the acceptable range print...

Page 25: ...well You can use the control solutions to verify the meter s functions Is the control solution expired or contaminated If yes replace with a new control solution to check the performance of SMBG syste...

Page 26: ...r you Please do not share your lancing device with anyone And always use a new sterile lancet Lancets are for one time use only Used test strips and lancets are considered bio hazardous waste in accor...

Page 27: ...cet into the lancing device to get it ready for use Lancets are for use only and a new sterile lancet should be used each time you perform a test 1 Unscrew the Cap 2 Insert the lancet into the lancing...

Page 28: ...lancing devices and lancets may transmit blood borne pathogens such as viral hepatitis 6 Set the lancing device aside until later in the test 5 Pull on the sliding barrel of the lancing device until i...

Page 29: ...g device against the pad of your finger Press the trigger button to activate the lancing device 2 Put a test strip into the meter in the direction of the arrow The meter turns on Press M button or C b...

Page 30: ...the result 8 Press power button to repeat the test result Or you can also press C button or M button to set the volume before repeating the result 9 Remove the test strip and the meter turns off auto...

Page 31: ...ervals In a pre meal or fasting state more than 2 hours since the last meal Two hours or more after taking insulin Two hours or more after exercise Do NOT use AST if You think your blood glucose is lo...

Page 32: ...sult from the alternate site test does not match how you feel do a fingertip test to confirm the result again Do NOT change your treatment just because of an alternate site result do a fingertip test...

Page 33: ...device for AST testing 1 Massage the puncture area of forearm for a few seconds 2 Press and hold the device with clear adjustable tip against the forearm 3 Press the trigger button to activate the la...

Page 34: ...conds 2 Press and hold the device with one clear adjustable tip against the palm 3 Press the trigger button to activate the lancing device 4 Hold the device against palm and increase pressure until th...

Page 35: ...he cap 2 Without touching the used lancet stick the lancet tip into its protective cover 3 Pointing the lancing device toward a container for sharp or biohazard material slide the ejection button down...

Page 36: ...on Standards of medical care in diabetes Diabetes care 2013 Vol 36 Suppl 1 S21 Unusual Test Results If your test result does not match the way you feel please follow these steps 1 Run a control test C...

Page 37: ...3 3 mmol L and you have symptoms of high blood glucose contact your healthcare professional instantly If the result does not match how you feel follow the steps under Unusual Test Results What It Mean...

Page 38: ...drawn from your arm by a healthcare professional Keep in mind that the lab could use different technology than EASYMAX Voice Blood Glucose Meter and that blood glucose meters for self testing generall...

Page 39: ...nge your therapy based on one individual result in memory 2 The memory is not lost when you replace the battery You do need to check that the time and date are still correct See Section Setting The Ti...

Page 40: ...individual test result 4 To delete a test result press C button for more than 3 seconds and display shows dEL Then press power button to confirm the deletion 5 Press C button to keep reviewing the res...

Page 41: ...ng the set mode you can personalize your meter to suit your needs Here are the features you can customize Time and Date to set the time and date see Chapter 1 Language to set the language see Chapter...

Page 42: ...ll not turn on the first time batteries are inserted Please press and hold Power Button or insert the test strip to turn your meter on The meter will turn off automatically Or you can press and hold P...

Page 43: ...cloth dampened with water and mild detergent DO NOT places the entire device under water To disinfect the cap after cleaning place it in 70 75 rubbing alcohol for 10 minutes at least once a week Allow...

Page 44: ...turn it on and lets you know if something is wrong See Screen Messages and what to do about them To make sure the display is working properly turn off the meter Press and hold Power Button to see the...

Page 45: ...local dealer or healthcare professional Message What it means What to do Humidified Used strips The meter has detected a problem with the test strip Repeat the test with a new strip Refer to pages 29...

Page 46: ...rs again please contact your local dealer Memory fault Replace the batteries first Refer to pages 16 and 42 If this error message appears again please contact your local dealer System fault There may...

Page 47: ...gain obtain and follow instructions from your healthcare professional without delay The test result is lower than 20 mg dL This may require immediate treatment according to your healthcare professiona...

Page 48: ...result may occur if the test continues Relocate the meter to a location with temperature between 10 C 40 C 50 F 104 F The Lt and thermometer icon appears Temperature is too low outside the required r...

Page 49: ...0 Storage and transportation condition Temp 20 C 50 C 4 F 122 F Relative Humidity R H 90 Blood sample 0 6 L Fresh blood from fingertip palm or forearm Hematocrit Hct 20 60 Power 2 Alkaline 1 5V AAA Ba...

Page 50: ...Hematocrit The hematocrit between 20 and 60 will not affect the results Hematocrit below 20 may cause higher results Hematocrit above 60 may cause lower results 8 Altitude up to 3 048 meters above sea...

Page 51: ...4 mmol L 250 mg dL 13 9 mmol L 500 mg dL 27 8 mmol L Ascorbic Acid 4 mg dL 0 26 mmol L 10 89 1 76 4 55 Ibuprofen 50 mg dL 2 43 mmol L 3 10 2 88 4 62 L Dopa 1 8 mg dL 0 09 mmol L 10 59 7 91 4 90 Sodium...

Page 52: ...ASYMAX US Office 9550 S Eastern Ave Suite 253 Las Vegas NV 89123 www easymaxdiabetescare com Obelis s a Boulevard G n ral Wahis 53 B 1030 Brussels Belgium TEL 32 2 732 59 54 FAX 32 2 732 60 03 E mail...

Reviews: