7 . Place bolt into the spacer as shown. Place knob onto
bolt and tighten securely.
8 . Remove pin from the pump, place pump handle onto
pump shaft as shown and place pin through holes. Place
top of handle on to the handle, insert 2 bolts and tighten
securely.
9 . Use the optional pump handle or step on the gas
spring assist pedal to raise the seat about 3/4 the way up
as shown. Line up the following arm with the back angle
tube.
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The EasyStand Bantam is intended for use in the electromagnetic environment specified below. The customer or the
user of the EasyStand Bantam should assure that it is used in such an environment.
Immunity Test
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80
MHz
3 Vrms
Portable and mobile RF communications equipment
should be used no closer to any part of the
EasyStand Bantam, including cables, than the rec-
ommended separation distance calculated from the
equation applicable to the frequency of the transmit-
ter.
Recommended separation distance
d = 1.2
√
P
d = 1.2
√
P
80 MHz to 800 MHz
d = 2.3
√
P
800 MHz to 2.5 GHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as deter-
mined by an electromagnetic site survey
a
, should
be less than the compliance level in each frequency
range.
b
Interference may occur in the vicinity of equipment
marked with the following symbol:
IEC 60601
Test Level
Electromagnetic Environment - Guidance
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
Compliance
Level
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5
GHz
3 V/m
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the EasyStand Bantam is used exceeds the
applicable RF compliance level above, the EasyStand Bantam should be observed to verify normal operation. If abnor-
mal performance is observed, additional measures may be necessary, such as re-orienting or relocating the
EasyStand Bantam.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Page 5