Unit description

Display information
Measuring location symbol:
 The icon indicates the correct 

measuring position.
Temperature data indicator: The temperature reading is 

displayed after measuring.
Temperature unit indicator: Unit of measuring temperature 

shows ˚F or ˚C.
Low battery symbol: The icon prompts you to replace the 

battery when the power is low.
Memory symbol: It shows in memory query mode currently.

Load and replace batteries
When the battery symbol “

” appears, please replace with new 

batteries immediately. The steps are as follows: 

1.  To remove the battery cover, press down on and slide the 

battery cover towards the bottom of the thermometer. 

2.  Install 2 x AAA batteries according to the correct  

positive and negative pole as shown on the inside  

of the battery compartment.

3.  Close the battery cover.

ON/OFF/SCAN

Unit body

Display screen

Probe

Intended use
Thank you for purchasing the EasyTouch® Infrared Thermometer, 

type AET-R1D1. This Infrared Thermometer is a non-sterile, 

reusable, handheld device. It can be used by consumers in 

homecare environment and doctors in clinic as reference. It 

is intended for measuring the human body temperature of 

individuals three months of age and older by detecting infrared 

heat from the center of the forehead.

•  For safe and proper use of this product, be sure to read  

and fully understand Safety Precautions contained in  

this user manual.

•  Keep this manual with you for a convenient reference.
•  Keep this manual in a proper way and avoid any loss.
•  If you need other information that is not listed in the 

instructions-for-use packet, please contact MHC Medical 

Products at (877) 358.4342.

The correct measurement method 

to follow: 
Measure forehead.  

Distance 0.59” to 1.97” (15mm to 50mm)

Correct measurement method

1.  Turn the thermometer on by 

pressing the “ON/OFF/SCAN”  

button. Choose mode by pressing  

“ON/OFF/SCAN” on the shutoff 

mode. A beep sound will be heard 

following which the screen will 

indication with ---˚F (˚C), indicating 

the unit is ready for use.

2.  Place the thermometer probe 

point near the forehead and press 

“ON/OFF/SCAN” and release after 

~1 second and the thermometer 

will beep and display the accurate 

temperature. Whenever consecutive 

readings are required, please wait 

for at least 5 seconds. The above 

procedure can be repeated as many 

times as required.

3.  To turn off the device press and  

hold the “ON/OFF/SCAN” button  

for 5 seconds. The thermometer will 

turn off automatically after  

60 seconds of not being in use.

1

2

3

Warning:
Please take out the battery if you do not use the thermometer 

for longer than a month. Please refer to the relevant laws and 

regulations for the disposal of waste batteries.

Before take a measurement
1.  
It is suggested to read through the entire instructions for use 

packet to understand how to operate the thermometer.

2.  In order to avoid affecting accuracy of the device due to the 

external too cold or too hot environment, it must be placed 

in general environment of room temperature of 60.8˚F-95˚F 

(16˚C-35˚C) for at least 30 minutes before use to avoid 

incorrect measurement results.

3.  The person measured should be kept at room temperature 

for at least 20 minutes to make body temperature balanced.

4.  After strenuous exercise, you should rest for at least 30 

minutes before taking measurement.

5.  Wipe the skin dry and push aside hair strands before 

commencing.

6.  When repeated measurements are made continuously,  

move away the probe between each measurement and  

rest for at least 5s before making the next to obtain the  

most correct result.

7.  The normal reading value is about 95.9˚F~100˚F 

(35.5˚C~37.8˚C). The user should frequently take his or her 

own temperature to know their average temperature. Pay 

attention to physical changes or consult your doctor if there 

is any abnormality. 

8.  Check to make sure the probe is clean before each 

measurement to ensure accuracy of temperature reading. 

See Page 7 for instructions on how to properly clean  

the thermometer.

Correct position to measure

Forehead temperature mode
The correct measurement method 

to follow:
Measure the forehead.
Distance 0.59” to 1.97” (15mm to 50mm)

Instructions for Use
1.  
Turn the thermometer on by pressing the  

“ON/OFF/SCAN” button.

2.  A beep sound will be then heard and the screen will then 

flash ----˚F (˚C). The unit is now ready for use. Using the 

correct measuring position, point the thermometer’s probe 

to the forehead, press “ON/OFF/SCAN” and release after 

~1 second, the thermometer will beep and display the 

temperature. If a consecutive reading is required then please 

wait for at least 5 seconds. Repeat the above procedure as 

often as required.

3.  Press and hold the “ON/OFF/SCAN” button for 5 seconds 

to turn off the thermometer. Or the unit will switch off 

automatically after 60 seconds.

•  If a reading below 89.6˚F (32˚C) is recorded a “Lo” sign will 

be displayed followed by 2 consecutive beeps.

•  If a reading above 100˚F (37.8˚C) is recorded there will be 

6 consecutive beeps.

•  If a reading above 107.96˚F (42.2˚C) is recorded a “Hi” sign 

will be displayed followed by 2 consecutive beeps.

Recalling Readings from Memory
1.  
Ensure the thermometer is switched off.

2.  Press and hold the “ON/OFF/SCAN” button for 4 seconds  

to put the device on memory mode. The last reading will 

be displayed.

3.  Press and release the “ON/OFF/SCAN” button again to 

display the next reading in a sequence from the last reading 

to the first reading.

4.  The thermometer has the memory to recall the last  

32 readings.

5.  The thermometer will automatically turn off if it idles for 

more than 12 seconds.

Changing Temperature Scales
1.  
Ensure the thermometer is switched off.

2.  Press and hold the “ON/OFF/SCAN” button for 8 seconds to 

bring it into temperature mode. The current temperature 

unit will be displayed.

3.  Release and press the “ON/OFF/SCAN” button again to 

toggle and select between ˚C and ˚F.

4.  The thermometer will switch off automaticaly if it idles for 

more than 4 seconds.

Disposal
Do not dispose of electrical appliances as unsorted 

municipal waste, use separate collection facilities. 

Contact you local government for information 

regarding the collection systems available. If electrical 

appliances are disposed of in landfills or dumps, 

hazardous substances can leak into the groundwater 

and get into the food chain, damaging your health and 

well-being.

Explanation of Marks or Symbols
The following symbols may appear on the manual, Infrared 

Thermometer and its accessories.

MUST OBSERVE 
Means Obligatory with detailed items expressed in 

words or figures within or beside the mark. Left one 

means General Compulsory.

Follow instructions for use.

IMPLICATION OF SYMBOL
Type-BF applied part.
CE Mark: conforms to essential requirements of the 

Medical Device Directive 93/42/EEC.

Caution: Consult accompanying documents.

Do not dispose this product as unsorted municipal 

waste. Collection of such waste separately for 

special treatment is necessary.

Transport package shall be kept away from rain.

Transport package shall not be exposed to sunlight.

Indicates correct upright position of the transport 

package.

Contents of the transport package are fragile 

therefore it shall be handled with care.

Indicates temperature limits within which the 

transport package shall be stored and handled.

Non-ionizing electromagnetic radiation

Do not roll

Lot number

Production date

The device should not be used after the expiration 

date listed on the retail box.
Protected against solid foreign objects of 12.5mm 

dianeter and greater. protected against vertically 

falling water drops when the device is tilted up 

to 15˚.

Manufacturer

Authorized Representative in the European 

Community

1

2

3

4

5

Memory symbol

Low battery 

symbol

Measuring location symbol

Temperature unit

indicator

Temperature data

indicator

1

2

3

Battery compartment

 

(enlarged to  

show detail)

Summary of Contents for AET-R1D1

Page 1: ...ons for use packet please contact MHC Medical Products at 877 358 4342 The correct measurement method to follow Measure forehead Distance 0 59 to 1 97 15mm to 50mm Correct measurement method 1 Turn the thermometer on by pressing the ON OFF SCAN button Choose mode by pressing ON OFF SCAN on the shutoff mode A beep sound will be heard following which the screen will indication with F C indicating th...

Page 2: ...ory query mode currently Load and replace batteries When the battery symbol appears please replace with new batteries immediately The steps are as follows 1 To remove the battery cover press down on and slide the battery cover towards the bottom of the thermometer 2 Install 2 x AAA batteries according to the correct positive and negative pole as shown on the inside of the battery compartment 3 Clo...

Page 3: ...e you should rest for at least 30 minutes before taking measurement 5 Wipe the skin dry and push aside hair strands before commencing 6 When repeated measurements are made continuously move away the probe between each measurement and rest for at least 5s before making the next to obtain the most correct result 7 The normal reading value is about 95 9 F 100 F 35 5 C 37 8 C The user should frequentl...

Page 4: ...r is switched off 2 Press and hold the ON OFF SCAN button for 4 seconds to put the device on memory mode The last reading will be displayed 3 Press and release the ON OFF SCAN button again to display the next reading in a sequence from the last reading to the first reading 4 The thermometer has the memory to recall the last 32 readings 5 The thermometer will automatically turn off if it idles for ...

Page 5: ...or special treatment is necessary Transport package shall be kept away from rain Transport package shall not be exposed to sunlight Indicates correct upright position of the transport package Contents of the transport package are fragile therefore it shall be handled with care Indicates temperature limits within which the transport package shall be stored and handled Non ionizing electromagnetic r...

Page 6: ...ns equipment can affect MEDICAL ELECTICAL EQUIPMENT Falling to do so may cause inaccuracy The probe glass is made from special material which is not easy broken If there is any damage stop using it and contact MHC Medical Products immediately Do not modify this equipment without authorization of the manufacturer Contact MHC Medical Products immediately The device is not repairable and contains no ...

Page 7: ...dy using a soft and dry cloth 2 Use a cotton swab or soft cloth absorbed with alcohol and gently wipe the lens of the temperature measuring head After using the product the product can be sterilized with 70 medical alcohol on the soft towel or cotton sliver If the probe tip is damaged please contact MHC Medical Products 3 Don t wash with water or detergent containing abrasive or benzene 4 Do not i...

Page 8: ...tically switches off after 1 minute Product life 5 years New battery life 1000 times Standard The product is made under the ISO80601 2 56 Guarantee Two year warranty is available from purchasing date excluding user caused failures listed below 1 Failure resulted in unauthorized disassembly and modification 2 Failure resulted in unexpected drop during application or transportation 3 Failure resulte...

Page 9: ...E If any the performance of the ME EQUIPMENT or ME SYSTEM that was determined to be ESSENTIAL PERFORMANCE and a description of what the OPERATOR can expect if the ESSENTIAL PERFORMANCE is lost or degraded due to EM DISTURBANCES the defined term ESSENTIAL PERFORMANCE need not be used 1 All necessary instructions for maintaining BASIC SAFETY and ESSENTIAL PERFORMANCE with regard to electromagnetic d...

Page 10: ...ectromagnetic Immunity Radiated RF IEC61000 4 3 Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment Test Frequency MHz Band MHz Service Modulation Modulation W Distance m IMMUNITY TEST LEVEL V m 385 380 390 TETRA 400 Pulse modulation 18 Hz 1 8 0 3 27 450 380 390 GMRS 460 FRS 460 FM 5 kHz deviation 1kHz sine 2 0 3 28 710 704 787 LTE Band 13 17 Pulse modulation 21...

Reviews: