600 A T-OP II™ terminator operation instructions
4
Operation Instructions
MN650047EN May 2017
Figure 11 . Open switch if it is closed
Reconnecting cable to apparatus bushings
Step 1
Determine that bushings are de-energized
●
●
Check apparatus switch and see that it is open. Open
switch if it is closed. (Refer to Figure 11.)
●
●
Grasp 600 A protective cap with shotgun. Rotate
counterclockwise, until threads are disengaged.
●
●
Remove protective cap from the bushing and set aside in
a clean, protected area.
●
●
Using, a volt meter, test apparatus bushing to determine
that the bushing is de-energized. (Refer to Figure 12.)
WARNING
Do not proceed until apparatus is de-energized .
Figure 12 . Test for voltage with volt meter
Step 2
Connect cable to apparatus bushing
●
●
Grasp pulling eye of the grounding elbow with shotgun,
pull firmly to remove elbow from the LRTP of the T-body.
●
●
Set elbow aside in a clean area.
●
●
Grasp O&T/Torque tool with the shotgun and pull fully
into hotstick.
●
●
Using the shotgun, insert the O&T/Torque tool into the
throat at the LRTP and engage the rotating nut. Rotate
tool counterclockwise until both the threaded connection
and the rubber interference fit is broken.
●
●
Lift the T-OP II termination from the insulated standoff
bushing to the apparatus bushing.
●
●
While pushing the T-body on to the apparatus bushing,
turn the O&T/Torque tool clockwise to make the threaded
connection. Turn until torque head rachets (20-25 ft-lbs.).
An audible click will be heard. (Refer to Figure 13.)
Figure 13 . Install T-OP II termination onto
apparatus bushing
●
●
Remove O&T/Torque tool.
●
●
Using the hotstick, replace 200 A protective cap or
elbow arrester on the 200 A interface of the LRTP.
(Refer to Figure 14.)