MGR10 A/B/C - Revision : I (for 7.0 software revision)
26
IMPORTANT NOTICES ON THE
/km and km MODES :
1)
The
/km
and
km
functions are not operating in RS232 or IEEE488 modes but are operating
in the PLC mode .
2)
The temperature compensation function stays operating .
3)
The comparison function with high and low limits , the analog output and the data storage are
not following the
/km
and
km
modes and continue to work on the real Rx value in Ohms .
4)
The ZERO function is not working in the
/km
and
km
modes
5)
Only the SLOW mode allows operating
/km
and
km
functions .
Trig Mode :
Press on [
Trig Mode
] to select the measurement trigger modes.
Press on [
Cont]
to activate the continuous measurement mode
Press on
[Sing ]
to activate single measurement mode. This mode performs a measurement
and memorizes the value on the display .To trigger a new measurement , press on the
[MES]
key .
Press on
[OK]
to go back to the previous page
Press on
[QUIT
] to escape the menu .
Read rate :
Press on
[Read rate ]
to select the measurement speed
Press on
[Slow]
to activate a display with large characters and with a rate of 2 measurements
per second which gives the best resolution and accuracy .
Press on
[Med]
to activate a display with large characters and with a rate of 4 measurements per
second which gives the best resolution and accuracy .
Or press on
[Fast ]
to perform measurement with a rate of 50 per second . This mode is
mainly advised when operating the unit from a remote interface . The measured values are
displayed in small characters and the accuracy is reduced .
Press on
[OK]
to go back to the previous page
Press on
[QUIT
] to escape the menu .
Filt :
Press on
[Filt]
to select the digital filter, this feature should be used when the sample to be
measured is unstable and the measured value displayed is constantly changing. The filter is a
rolling average of n measurements, where n may be set between 1 and 32.
Press on
[No. Rdgs]
to enter the number of measurements :
No of readings : 10
.
Press on [
Chg
] to change the value and enter the value with the numeric keypad .
Press on [
OK
] to enter the new value .This value will be displayed in the right upper
corner of the LCD screen if the filter is activated .
Filt : n
Or press on [
OK
] to keep the previous value
Press on [
On
] to activate the filter mode.