MGR10 A/B/C - Revision : I (for 7.0 software revision)
57
6.2.2.3 Interface Functions
The interface performs the following functions;
SH1, AH1, T5, TE0, L3, LE0, SR1, RL1, PP0, DC1, DT1, C0, E2
These have the following meanings :
SH1 Source
Handshake
The device can send multiline messages.
AH1 Acceptor
Handshake
The device can receive multiline messages.
T5 Talker
Capability
The device;
can act as a basic talker
;
can respond to a serial poll
; can perform a talk only mode
; unaddresses if MLA sent
TE0 Extended
Talker
Capability
Extended talker addressing is not implemented.
L3 Listener
Capability
The device;
can act as a basic listener
; can be set to listen only mode
; becomes unaddressed if MTA sent
LE0 Extended
Listener
Capability
Extended listener addressing is not implemented
SR1 Service
Request
Function
The device can asynchronously assert the SRQ line to request the attention of the system
controller.
RL1 Remote/Local
Function
The device has full remote/local switching capability. There is no manual switch on the
instrument.
PP0 Parallel
Poll
Function
The parallel poll function is not implemented.
DC1 Device Clear Function
The device may be set to the default state by the appropriate command, and includes a selective
device clear.
DT1 Device Trigger Function