7. 

Drying of hearing aids  

Place the hearing aid (max. 2 hearing aids) on the round rubber pad and 
put the drying cup over it.Before doing so it is recommended to remove 
the hearing aid  ear hooks and /or filter systems and to open the battery 
flap. The hearing system should 

not 

be placed directly above the vacuum 

opening (small opening in the middle) in ordert to create an optimal 
vacuum (-0,8 bar).  

TIP Please press the drying cup firmly onto the rubber pad for a moment, 
to ensre a faster pressure build-up. 

 

By pressing the left orange button drying , the five minutes drying display 
mode is activated. The segment display shows 5 minutes and counts 
them down. Fort he last 30seconds the 1 minute display starts blinking. 
The device is then automatically switchd off . Another pressing of the 
button drying means 15 minutes long term drying. After that the device 
switches off automatically.Five minutes drying starts again as soon as 
the left button drying is pressed again. To switch off the device the button 
STOP  must be pressed for at least 2 seconds. 
 

8. 

Care and cleaning e - cleaner

 

We recommend to clean the filter of  the vacuum pump every day (but at 
least once a week). It is located at the back of the device . In case of 
non-observance the device may be damaged. 

Remove the filter housing 

from the holder for cleaning, unplug  transparent hoses right and left, turn 
the cover oft he housing 

on the right side 

by 30 degrees counter 

clockwise. (Before doing so slide unlocker to the left).  

 
The moving part of the filter housing must be on the right side in order to 
collect the ear wax inside the filter. 

Remove cover (before doing so slide 

unlocker to the left ) Turn housing and empty the filter over a paper 
tissue. The filter must not come into contact with moisture. Remove filter 
from the housing. Ear wax is settled inside the filter. We recommend to 
replace the filter of the vacuum pump every 6 months against a new 
one.Replacement filter is included. Further filters available in the eccola 
shop under accessories.  
 
 
 
 

Summary of Contents for e-cleaner

Page 1: ..._________________________________________ Page 1 of 12 Status 04 2022 Operating Instructions for Drying and Cleaning Device for hearing aids with UV C desinfection e cleaner e cleaner optionally with...

Page 2: ...ilters 1 x Drying cup 1 x Dry Cap UV3 1 x Rubber pad self adhesive for Dry Cap UV3 1 x ecco clean cleaning liquid 1 x Operating instructions incl safety data sheet Congratulations On your purchase of...

Page 3: ...ning e cleaner 5 5 Cleaning of receiver and microphone of hearing aids 6 6 Short term hygiene program wiht Dry Cap UV3 UV C 6 7 Drying of hearing aids 7 8 Care and cleaning of e cleaner 7 9 Service in...

Page 4: ...ransportation Regulations Installation Instructions The e cleaner is delivered with power supply unit 12 V DC and input voltage of 110 V 240 V 60 50 Hz 0 25 A output 12 V DC 1000mA 12 VA The ambient t...

Page 5: ...urpose see photo Place the Dry Cap 3 on the rubber pad Photo rubber pad Dry Cap UV3 with suction needles and holder optionally and insert the USB cable on the back into the USB port of the e cleaner a...

Page 6: ...he reciever with ecco cleaning liquid 1 bottle included in the delivery and can be reordered Leave it to take effect for some time and clean the reciver again with the red vacuum hose If you press the...

Page 7: ...itches off automatically Five minutes drying starts again as soon as the left button drying is pressed again To switch off the device the button STOP must be pressed for at least 2 seconds 8 Care and...

Page 8: ...aced correctly the cover has been put on incorrectly or the hoses have not been inserted correctly Please repeat the whole process From time to time it may happen that the cleaning attachments are dir...

Page 9: ...forget to replace the filter Responsibilty for replacing the filter after this interval is on the user s side b After 750 h about 3 years the complete e cleaner should be sent to us for a complete se...

Page 10: ...e cleaner to be checked 12 Accessories Spare Parts You will find accessories and spare parts in our online shop www eccola de such as suction needle set suction needle holders red spare hose with hand...

Page 11: ...g directives EMC directive 2014 65 EC EN 55014 2 2016 01 EN 61000 6 1 2007 10 In the event of unautorized changes improper repairs or modifications this declaration loses its validity unless explicitl...

Page 12: ..._______________________________________________________________ Page 12 of 12 Status 04 2022 16 Sicherheitsdatenblatt ecco clean Reinigungsl sung...

Reviews: