Troubleshooting
6-34
Algas-SDI Operations & Maintenance Manual P/N 52632
Adjustment of Operating Temperature
WARNING
Dangerous voltages are present during this test. Use extreme
caution.
SETUP CONDITIONS
1. Turn the vaporizer power on.
2. Do not start the vaporizer.
3. Set the meter to DC volts and put the negative lead on TP7.
“OFF” TEMPERATURE ADJUTSMENT
To adjust the "
OFF"
temperature, put the positive meter lead on
TP3 and adjust R7 for 0.95 VDC ( 95°C ).
This reading must always be higher than the reading at TP4.
NOTE
Clockwise rotation increases temperature setting.
“ON” TEMPERATURE ADJUSTMENT
To adjust "
ON"
temperature, put the positive meter lead on TP4
and adjust R8 for 0.90 VDC ( 90°C ).
CORE TEMPERATURE MONITORING
Put the positive meter lead on TP1 and turn the vaporizer on.
The heaters should turn off and on at the set temperatures as
measured here. Observe the heat "
ON"
lamp on the control
board or on the control box.