3.In the switch on state,
Press the “°C/ °F” button to
transfer between Celsius and Fahrenheit.
4.In the switch on state,
Press “MEM” (Memory) button,
which will then display the last temperature, and
allows for a view of the last 32 measurements.
In the switch on state,Press "MEM" button and hold
for 5 seconds, all data in memory will be deleted. Then
press "MEM" button again, the display will show "CLr"
5.
In the switch on state, press the “LANGUAGE”
button will display a choice of three languages,
and an “OFF” setting, which will be silence the
talking feature of the unit.
6.Recalibration of device via the F4 MENU
When there is the difference between EM-Th1 and
mercury thermometer, and you believe mercury
thermometer from its temperature but it is not
convenient to use. You can use recalibration function
to adjust the EM-Th1 to make it the same test result
with mercury thermometer after recalibration.
Besides, when you use EM-Th1 for the people with
different skin color (For example: the yellow race, the
white race, black people and so on) you can use
recalibration too.
Instructions for recalibration:
In the switch on state, Press the “MODE” and “°C/ °F”
buttons at the same time for 3 seconds.
The screen will then display : “F4”
Press “MODE” to increses 0.1°F, press °C/ °F to
reduce 0.1°F.
Press “MEM” button to save the setting.
In the cases of seasonal or environmental changes a
verification and adjustment should be carried out.
7.Changing The Batteries
Display: when the LCD screen displays the flashed
symbol " ", the battery is used.
XV. EXPLANATION OF SYMBOLS
Symbol
Reference
Trade mark
IEC 60417-5333,Type BF applied part
IEC 60417-5031 Direct current
Protected against access to hazardous parts with a
finger and against vertically falling water drops when
enclosure tilted up to 15°
Refer to instruction manual / booklet
DISPOSAL: Do not dispose this product as unsorted
municipal waste. Collection of such waste separately
for special treatment is necessary.
Specifies serial number
SN