Range 3030
– 3031ARD
Instruction manual
ECOSAFE | Date of the document : 26/06/2019 – Version : 01
8
Connection to a ventilation system
The air inlet and outlet are situated in the back of the cabinet. See the below outline.
How to choose a ventilation or a filtration box ?
Ventilation boxes
are connected to the safety cabinet with a connecting kit (including a connecting sleeve, 2 hose clamps and 1 meter
of flexible sheath) ; the fan included in the box extracts the toxic vapors contained in the cabinet, and reject it to the outside of the
building thanks to the roof or wall extraction.
If you cannot reject the vapors to the outside, you need a filtration box.
Filtration boxes
are equipped with an activated carbon filter, which purifies the toxic vapors in the box, and rejects it in the room in a
healthy fashion. Filtration boxes are delivered with the activated carbon filter that you need to renew once a year.
Running without any mechanic ventilation
It is possible for your safety cabinet to run without mechanical ventilation. However, your cabinet has to be placed in a location that
provides its own ventilation system (at least 5 times the volume of the location, per hour).
The cabinet itself and a radius of 2,5 meters directly around is a dangerous area subject to explosions.
On the other hand, if the ventilation system of your cabinet is not running, you have to indicate it, by putting an etiquette on the cabinet,
saying “Safety cabinet not connected to the mechanic ventilation” (etiquette is not provided).
7. SAFETY CONTROLS
Your safety cabinet is a safety equipment. As to guarantee its perfect efficiency, you have to check the equipment regularly.