Range 4
– B04M
Operating instructions
ECOSAFE | Date du document : 30/04/2019 – Version : 01
3
40°C
Transport and handling
The safety cabinet that you received has been protected and transported in optimal conditions, as to be protected from any damage.
It has to be lifted only by suitable equipment (suct
ion cup, strap…). It has to be carried in a vertical position, and it must not be lying.
It is imperative not to tilt the cabinet when it is carried. Indeed it can cause damages that can alter the right functioning of the cabinet.
On the other hand, if it has to be moved, the cabinet has to be placed on a pallet and strapped, as to avoid any fall.
Commissioning
After having unwrapped the cabinet, ensure that you also received a set of two keys.
Your safety cabinet has to be installed, used and maintained as to guarantee the users protection.
You need to remove the covers on the vents before storing any can in your cabinet. To do this, simply unscrew them counterclockwise.
The doors open by pulling the handle towards you.
4. INTERNAL EQUIPMENTS
The cabinet is provided with 1 shelf. This must be fixed at the chosen height thanks to the rack and pinion.
Take care not to exceed the loading capacity per shelf of 75 KG.
The cabinet also has an integrated retention tank, which has a retention capacity of 6 liters.
The cabinet is provided with polyethylene trays attached to the galvanized steel shelves and a separate polyethylene liner for the
bottom sump.
5. STORAGE
Your safety cabinet is exclusively destined to store chemicals in laboratories, industries, schools
… in closed containers.
It must not store high concentrated corrosive products.
We remind you that the storage of corrosives can alter the quality of your safety equipment. On the other hand, we remind you that
damages due to a bad use of the equipment won
’t be covered by our warranty.
6. VENTILATION
We highly recommend connecting your safety cabinet to a ventilation system, as to evacuate the eventual vapors contained in it.
The air renewal has to be at least 10 times the volume of the cabinet, per hour. Moreover, ensure that the ventilation runs 24h/7.
Connection to a ventilation system
The air inlet and outlet are situated on the side of the cabinet.
How to choose a ventilation or a filtration box ?
Ventilation boxes
are connected to the safety cabinet with a connecting kit (including a connecting sleeve, 2 hose clamps and 1 meter
of flexible sheath) ; the fan included in the box extracts the toxic vapors contained in the cabinet, and reject it to the outside of the