TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Report on the performance testing of the Serinus 30 ambient air quality moni-
toring system manufactured by Ecotech Pty Ltd measuring CO, Report no.:
936/21221977/D_EN
Page 103 of 253
EP_Ecotech_Serinus30_en.docx
7.6 Detailed presentation of test results
Individual results are provided in Table
Table 26:
Individual results for the difference between sample and calibration port
Time
Device 1 [µmol/mol] Device 2 [µmol/mol]
17:50
64.8
64.3
17:51
64.8
64.3
17:52
64.8
64.3
Average
64.8
64.3
17:59
64.9
64.4
18:00
64.9
64.4
18:01
64.9
64.4
Average
64.9
64.4
Deviation [%]
-0.09
-0.17
Test gas to the sample port
Test gas to the calibration port