TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Seite 16 von 253
Report on the performance testing of the Serinus 30 ambient air quality
monitoring system manufactured by Ecotech Pty Ltd measuring CO, Report
no.: 936/21221977/D_EN
EP_Ecotech_Serinus30_en.docx
Performance criterion
Minimum requirement
Test result
Com-
pliance
Page
The sensitivity coefficient of
sample gas temperature at ref-
erence point shall not exceed
the requirements listed in Ta-
ble 2 of VDI Guideline 4202
(September 2010).
Please refer to section 7.1 8.4.8
Sensitivity coefficient to sample
The sensitivity coefficient of
surrounding temperature at ze-
ro and reference point shall not
exceed the requirements listed
in Table 2 of VDI Guideline
4202 Sheet 1 (September
2010).
Please refer to section 7.1 8.4.9
Sensitivity coefficient to the sur-
The sensitivity coefficient of
supply voltage shall not exceed
the requirements listed in Table
2 of VDI Guideline 4202 Sheet
1 (September 2010).
8.4.10 Sensitivity coefficient to
The change in the measured
value caused by interfering
components in the sample gas
shall not exceed the require-
ments listed in Table 2 of VDI
Guideline 4202 Sheet 1 (Sep-
tember 2010) at zero and ref-
erence point.
Summary of Contents for Serinus 30
Page 120: ...Serinus 30 Kohlenmonoxid Analysator Benutzerhandbuch Version 2 1 www ecotech com...
Page 121: ...Serinus 30 Benutzerhandbuch 2 1 Seite 2 Diese Seite wurde absichtlich frei gelassen...
Page 135: ...Serinus 30 Benutzerhandbuch 2 1 Seite 16 Diese Seite wurde absichtlich frei gelassen...
Page 181: ...Serinus 30 Benutzerhandbuch 2 1 Seite 62 Diese Seite wurde absichtlich frei gelassen...
Page 197: ...Serinus 30 Benutzerhandbuch 2 1 Seite 78 Diese Seite wurde absichtlich frei gelassen...
Page 229: ...Serinus 30 Benutzerhandbuch 2 1 Seite 110 Diese Seite wurde absichtlich frei gelassen...
Page 237: ...Serinus 30 Benutzerhandbuch 2 1 Seite 118 Diese Seite wurde absichtlich frei gelassen...