Report on the performance testing of the Serinus 30 ambient air quality
monitoring system manufactured by Ecotech Pty Ltd measuring CO, Report
no.: 936/21221977/D_EN
EP_Ecotech_Serinus30_en.docx
where
I
is the light intensity measured with CO in the gas sample
I
0
is the light intensity measured with no ozone in the gas sample
a
is the CO absorption coefficient at 253.7 nm
c
is the mass concentration of CO in mg/m
3
d
is the optical path length in m
Figure 2:
Illustration of the theoretical measurement method within the cell
The system includes a gas filter correlation wheel. It contains three parts which improve the
accuracy of the measurements: CO, N
2
and a mask.
The CO cell contains a particular saturation of CO (40 %) and thus helps to create a ref-
erence beam as it absorbs a known amount of light.
The N
2
cell contains 100% N
2
. It does not absorb IR radiation at a wavelength of 4.7 µm
and is therefore used during the normal measurement of CO.
The cover completely blocks the light source. It is used to determine background signals
in relation to each other and in relation to the background.
Summary of Contents for Serinus 30
Page 120: ...Serinus 30 Kohlenmonoxid Analysator Benutzerhandbuch Version 2 1 www ecotech com...
Page 121: ...Serinus 30 Benutzerhandbuch 2 1 Seite 2 Diese Seite wurde absichtlich frei gelassen...
Page 135: ...Serinus 30 Benutzerhandbuch 2 1 Seite 16 Diese Seite wurde absichtlich frei gelassen...
Page 181: ...Serinus 30 Benutzerhandbuch 2 1 Seite 62 Diese Seite wurde absichtlich frei gelassen...
Page 197: ...Serinus 30 Benutzerhandbuch 2 1 Seite 78 Diese Seite wurde absichtlich frei gelassen...
Page 229: ...Serinus 30 Benutzerhandbuch 2 1 Seite 110 Diese Seite wurde absichtlich frei gelassen...
Page 237: ...Serinus 30 Benutzerhandbuch 2 1 Seite 118 Diese Seite wurde absichtlich frei gelassen...