TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Report on the performance testing of the Serinus 30 ambient air quality moni-
toring system manufactured by Ecotech Pty Ltd measuring CO, Report no.:
936/21221977/D_EN
Page 73 of 253
EP_Ecotech_Serinus30_en.docx
7.4
Evaluation
Table 5:
Response times of the two Serinus 30 measuring systems for CO
For system 1 this results in a maximum t
r
of 50 s, a maximum t
f
of 52 s and a t
d
of -2 s for
CO.
For system 2 this results in a maximum t
r
of 51 s, a maximum t
f
of 56 s and a t
d
of -5 s for
CO.
7.5
Assessment
The maximum permissible response time of 180 s is exceeded at no time. The maximum re-
sponse time determined is 52 s for system 1 and 56 s for system 2.
Does this comply with the performance criterion? yes
Requirement
≤ 180 s
50
51
≤ 180 s
52
56
≤ 10 s
-2
-5
Average rise t
r
[s]
Average fall t
f
[s]
Difference t
d
[s]
Device 1
Device 2
Summary of Contents for Serinus 30
Page 120: ...Serinus 30 Kohlenmonoxid Analysator Benutzerhandbuch Version 2 1 www ecotech com...
Page 121: ...Serinus 30 Benutzerhandbuch 2 1 Seite 2 Diese Seite wurde absichtlich frei gelassen...
Page 135: ...Serinus 30 Benutzerhandbuch 2 1 Seite 16 Diese Seite wurde absichtlich frei gelassen...
Page 181: ...Serinus 30 Benutzerhandbuch 2 1 Seite 62 Diese Seite wurde absichtlich frei gelassen...
Page 197: ...Serinus 30 Benutzerhandbuch 2 1 Seite 78 Diese Seite wurde absichtlich frei gelassen...
Page 229: ...Serinus 30 Benutzerhandbuch 2 1 Seite 110 Diese Seite wurde absichtlich frei gelassen...
Page 237: ...Serinus 30 Benutzerhandbuch 2 1 Seite 118 Diese Seite wurde absichtlich frei gelassen...