TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Report on the performance testing of the Serinus 30 ambient air quality moni-
toring system manufactured by Ecotech Pty Ltd measuring CO, Report no.:
936/21221977/D_EN
Page 81 of 253
EP_Ecotech_Serinus30_en.docx
Table 11:
Individual test results for the repeatability standard deviation
Device 1
Device 2
Device 1
Device 2
Time
[µmol/mol]
[µmol/mol]
Time
[µmol/mol]
[µmol/mol]
13:05:01
-0.1
0.1
12:32:00
9.2
9.6
13:05:56
-0.1
0.1
12:32:55
9.2
9.6
13:06:51
-0.1
0.1
12:33:50
9.2
9.6
13:07:46
-0.1
0.1
12:34:45
9.2
9.6
13:08:41
-0.1
0.1
12:35:40
9.2
9.6
13:09:36
0.0
0.1
12:36:35
9.2
9.6
13:10:31
0.0
0.1
12:37:30
9.2
9.6
13:11:26
0.0
0.1
12:38:25
9.2
9.6
13:12:21
0.0
0.1
12:39:20
9.2
9.6
13:13:16
0.0
0.1
12:40:15
9.1
9.6
13:14:11
0.0
0.1
12:41:10
9.2
9.6
13:15:06
0.0
0.1
12:42:05
9.1
9.6
13:16:01
0.0
0.1
12:43:00
9.1
9.6
13:16:56
0.0
0.1
12:43:55
9.1
9.6
13:17:51
0.0
0.1
12:44:50
9.1
9.6
13:18:46
0.0
0.1
12:45:45
9.1
9.6
13:19:41
0.0
0.1
12:46:40
9.1
9.6
13:20:36
0.0
0.1
12:47:35
9.1
9.6
13:21:31
0.0
0.1
12:48:30
9.1
9.6
13:22:26
0.0
0.1
12:49:25
9.1
9.6
Average
0.0
0.1
Average
9.1
9.6
Zero level
c
t
level
Summary of Contents for Serinus 30
Page 120: ...Serinus 30 Kohlenmonoxid Analysator Benutzerhandbuch Version 2 1 www ecotech com...
Page 121: ...Serinus 30 Benutzerhandbuch 2 1 Seite 2 Diese Seite wurde absichtlich frei gelassen...
Page 135: ...Serinus 30 Benutzerhandbuch 2 1 Seite 16 Diese Seite wurde absichtlich frei gelassen...
Page 181: ...Serinus 30 Benutzerhandbuch 2 1 Seite 62 Diese Seite wurde absichtlich frei gelassen...
Page 197: ...Serinus 30 Benutzerhandbuch 2 1 Seite 78 Diese Seite wurde absichtlich frei gelassen...
Page 229: ...Serinus 30 Benutzerhandbuch 2 1 Seite 110 Diese Seite wurde absichtlich frei gelassen...
Page 237: ...Serinus 30 Benutzerhandbuch 2 1 Seite 118 Diese Seite wurde absichtlich frei gelassen...