TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Seite 92 von 253
Report on the performance testing of the Serinus 30 ambient air quality
monitoring system manufactured by Ecotech Pty Ltd measuring CO, Report
no.: 936/21221977/D_EN
EP_Ecotech_Serinus30_en.docx
7.3
Testing
The test was performed in accordance with the requirements on testing as stipulated in DIN
EN 14626.
7.4
Evaluation
The following sensitivity coefficients to the surrounding temperature resulted from the tests.
Table 18:
Sensitivity coefficient to the surrounding temperature at zero point and at span point,
systems 1 and 2
As illustrated in Table 18, the sensitivity coefficient to the surrounding temperature at zero
point and at span point complies with the performance criteria.
Requirement
≤ 0.3
0.009
0.035
≤ 0.3
0.005
0.013
≤ 0.3
0.040
0.046
≤ 0.3
0.029
0.008
Sensitivity coefficient at 30 ºC for span level [µmol/mol/K]
Device 1
Device 2
Sensitivity coefficient at 0 ºC for zero level [µmol/mol/K]
Sensitivity coefficient at 30 ºC for zero level [µmol/mol/K]
Sensitivity coefficient at 0 ºC for span level [µmol/mol/K]
Summary of Contents for Serinus 30
Page 120: ...Serinus 30 Kohlenmonoxid Analysator Benutzerhandbuch Version 2 1 www ecotech com...
Page 121: ...Serinus 30 Benutzerhandbuch 2 1 Seite 2 Diese Seite wurde absichtlich frei gelassen...
Page 135: ...Serinus 30 Benutzerhandbuch 2 1 Seite 16 Diese Seite wurde absichtlich frei gelassen...
Page 181: ...Serinus 30 Benutzerhandbuch 2 1 Seite 62 Diese Seite wurde absichtlich frei gelassen...
Page 197: ...Serinus 30 Benutzerhandbuch 2 1 Seite 78 Diese Seite wurde absichtlich frei gelassen...
Page 229: ...Serinus 30 Benutzerhandbuch 2 1 Seite 110 Diese Seite wurde absichtlich frei gelassen...
Page 237: ...Serinus 30 Benutzerhandbuch 2 1 Seite 118 Diese Seite wurde absichtlich frei gelassen...