TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Seite 22 von 253
Report on the performance testing of the Serinus 30 ambient air quality
monitoring system manufactured by Ecotech Pty Ltd measuring CO, Report
no.: 936/21221977/D_EN
EP_Ecotech_Serinus30_en.docx
2.
Task definition
2.1 Nature of the test
Ecotech Pty Ltd has commissioned TÜV Rheinland Energie und Umwelt GmbH to carry out
performance testing of the Serinus 30 measuring system. The test was a complete perfor-
mance test.
2.2 Objective
The instrument is designed to measure the concentration of CO in ambient air in the follow-
ing concentration range:
Component
Certification range
Unit
Carbon monoxide
0 - 100
mg/m³
The Serinus 30 uses the method of non-dispersive infrared spectroscopy to measure CO.
Performance testing was to be carried out in accordance with current standards taking into
consideration the latest developments.
Testing was performed on the basis of the following standards and guidelines:
•
VDI Guideline 4202 Sheet 1: Performance criteria for performance tests of automated
ambient air measuring systems; Point-related measurement methods for gaseous
and particulate air pollutants, September 2010
•
VDI Guideline 4203 Sheet 3: Testing of automated measuring systems; Test proce-
dures for point related ambient air measuring systems for gaseous and particulate air
pollutants, September 2010
•
DIN EN 14626: Ambient air – Standard method for the measurement of the concen-
tration of carbon monoxide by non-dispersive infrared spectroscopy, December 2012
Summary of Contents for Serinus 30
Page 120: ...Serinus 30 Kohlenmonoxid Analysator Benutzerhandbuch Version 2 1 www ecotech com...
Page 121: ...Serinus 30 Benutzerhandbuch 2 1 Seite 2 Diese Seite wurde absichtlich frei gelassen...
Page 135: ...Serinus 30 Benutzerhandbuch 2 1 Seite 16 Diese Seite wurde absichtlich frei gelassen...
Page 181: ...Serinus 30 Benutzerhandbuch 2 1 Seite 62 Diese Seite wurde absichtlich frei gelassen...
Page 197: ...Serinus 30 Benutzerhandbuch 2 1 Seite 78 Diese Seite wurde absichtlich frei gelassen...
Page 229: ...Serinus 30 Benutzerhandbuch 2 1 Seite 110 Diese Seite wurde absichtlich frei gelassen...
Page 237: ...Serinus 30 Benutzerhandbuch 2 1 Seite 118 Diese Seite wurde absichtlich frei gelassen...