TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Report on the performance testing of the Serinus 30 ambient air quality moni-
toring system manufactured by Ecotech Pty Ltd measuring CO, Report no.:
936/21221977/D_EN
Page 33 of 253
EP_Ecotech_Serinus30_en.docx
4.
Test programme
4.1 General
Performance testing was carried out with two complete and identical instruments with the se-
rial number
System 1: SN 12-1183 and
System 2: SN 13-0093.
During the test software version 2.09.0005 was implemented.
Performance testing consisted of a laboratory test to determine the performance characteris-
tics and a field test that lasted several months.
This report presents a heading for each test criterion along with the number and description
as stipulated in the respective standard [1. 2. 3. 4].
4.2 Laboratory test
The laboratory test was carried out with two identical Serinus 30 measuring systems with the
serial numbers SN: 12-1183 und SN: 13-0093. In accord with the guidelines [2, 3] the follow-
ing performance criteria were tested in the laboratory:
•
Description of operating states
•
General requirements
•
Adjustment of the calibration line
•
Short-term drift
•
Repeatability standard deviation
•
Sensitivity coefficient of the sample gas pressure
•
Sensitivity coefficient of surrounding temperature
•
Sensitivity coefficient of supply voltage
•
Cross-sensitivities
•
Response time
•
Difference between sample and calibration port
Instrument readings were recorded using an external data logger.
Results obtained during the laboratory tests are summarised in section 6.
Summary of Contents for Serinus 30
Page 120: ...Serinus 30 Kohlenmonoxid Analysator Benutzerhandbuch Version 2 1 www ecotech com...
Page 121: ...Serinus 30 Benutzerhandbuch 2 1 Seite 2 Diese Seite wurde absichtlich frei gelassen...
Page 135: ...Serinus 30 Benutzerhandbuch 2 1 Seite 16 Diese Seite wurde absichtlich frei gelassen...
Page 181: ...Serinus 30 Benutzerhandbuch 2 1 Seite 62 Diese Seite wurde absichtlich frei gelassen...
Page 197: ...Serinus 30 Benutzerhandbuch 2 1 Seite 78 Diese Seite wurde absichtlich frei gelassen...
Page 229: ...Serinus 30 Benutzerhandbuch 2 1 Seite 110 Diese Seite wurde absichtlich frei gelassen...
Page 237: ...Serinus 30 Benutzerhandbuch 2 1 Seite 118 Diese Seite wurde absichtlich frei gelassen...