TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Report on the performance testing of the Serinus 30 ambient air quality moni-
toring system manufactured by Ecotech Pty Ltd measuring CO, Report no.:
936/21221977/D_EN
Page 75 of 253
EP_Ecotech_Serinus30_en.docx
7.1
8.4.4 Short-term drift
The short-term drift at zero shall not exceed 0.10 µmol/mol/12h (i.e.
0.12 mg/m³/12h).
The short-term drift at span level shall not exceed 0.60 µmol/mol/12h (i.e. 0.70
mg/m³/12h).
7.2
Test procedure
After the required stabilisation period, the analyser shall be adjusted at zero and span level
(around 70 % to 80 % of the maximum of the certification range of carbon monoxide). After
waiting the time equivalent to one independent reading, 20 individual measurements are rec-
orded, first at zero and then at span concentration. From these 20 measurements, the aver-
age is calculated for zero and span level.
The analyser shall be kept running under the laboratory conditions while analysing ambient
air. After a period of 12 h, zero and span gas is fed to the analyser. After waiting the time
equivalent to one independent reading, 20 individual measurements are recorded, first at ze-
ro and then at span concentration. The averages for zero and span level shall be calculated.
The short-term drift at zero and span level shall be calculated as follows:
)
(
1
,
2
,
,
Z
Z
Z
S
C
C
D
−
=
where
Z
S
D
,
is the 12-hour drift at zero, in µmol/mol;
1
,
Z
C
is the average concentration of the measurements at zero at the beginning of the
drift period, in µmol/mol;
2
,
Z
C
is the average concentration of the measurements at zero at the end of the drift pe-
riod, in µmol/mol.
Z
S
D
,
shall comply with the performance criterion as specified above.
Z
S
S
S
S
S
D
C
C
D
,
1
,
2
,
,
)
(
−
−
=
where
S
S
D
,
is the 12-hour drift at span level, in µmol/mol;
1
,
S
C
is the average concentration of the measurements at span level at the beginning of
the drift period, in µmol/mol;
2
,
S
C
is the average concentration of the measurements at span level at the end of the
drift period, in µmol/mol.
S
S
D
,
shall comply with the performance criterion as specified above.
Summary of Contents for Serinus 30
Page 120: ...Serinus 30 Kohlenmonoxid Analysator Benutzerhandbuch Version 2 1 www ecotech com...
Page 121: ...Serinus 30 Benutzerhandbuch 2 1 Seite 2 Diese Seite wurde absichtlich frei gelassen...
Page 135: ...Serinus 30 Benutzerhandbuch 2 1 Seite 16 Diese Seite wurde absichtlich frei gelassen...
Page 181: ...Serinus 30 Benutzerhandbuch 2 1 Seite 62 Diese Seite wurde absichtlich frei gelassen...
Page 197: ...Serinus 30 Benutzerhandbuch 2 1 Seite 78 Diese Seite wurde absichtlich frei gelassen...
Page 229: ...Serinus 30 Benutzerhandbuch 2 1 Seite 110 Diese Seite wurde absichtlich frei gelassen...
Page 237: ...Serinus 30 Benutzerhandbuch 2 1 Seite 118 Diese Seite wurde absichtlich frei gelassen...