Service
Page 119
13.
Inspect the instrument’s plumbing looking for obvious damage. Check the condition of fittings,
particulate filter housing plus the O rings both in the filter assembly and in the cell assembly.
14.
If the leak is still present divide up the pneumatic system into discrete sections to locate the leak
(refer to Section 9.5). When the location of the leak has been determined repair and then rerun
the leak check procedure.
15.
Open -
Main Menu
Service Menu
Diagnostics Menu
Valve Menu
.
16.
Toggle Off –
Sample/Cal
Off
(this is done to release the vacuum safely)
17.
Allow one minute for the system to equalise even when your gauge is back at ambient (residual
vacuum will still be in sections of the system. Remove the blocker on the
BGnd Air Port
and
Inspect the internal tubing to ensuring that the tubing is cleanly connected to the fittings and
that the internal Teflon lining has not been kinked or crumpled.
18.
Remove the leak check jig.
19.
Enable -
Valve Sequencing
Enabled
.
6.3.6
Replacing the PMT Desiccant Pack
The PMT housing contains two desiccant packs to prevent condensation on the PMT cold block
housing. If the desiccant expires it will result in corrosion and premature cooler failure. It is
recommended that the desiccant bags be changed at least annually. If moisture is detected inside
the housing or the desiccant packs are saturated the interval should be reduced. To change the
desiccant packs follow the instructions below:
CAUTION
Because the PMT is extremely sensitive to light, it is essential that before opening
the PMT assembly to make sure that the instrument is switched off.
Once the instrument is switched off it is very important to cover the PMT at all
times so that no direct light reaches its window.
Equipment Required
Phillips Head Screwdriver
New Desiccant Packs
Tweezers
Procedure
1.
Turn the instrument off and disconnect the power. Wait 15 min for the cold block to warm up.
2.
Using a Phillips head screwdriver, remove the desiccant pack access cap from the PMT housing.
Summary of Contents for Serinus 44
Page 1: ...Serinus 44 Ammonia Oxides of Nitrogen Analyser User Manual Version 3 0 www ecotech com...
Page 2: ...Serinus 44 User Manual 3 0 Page 2 This page is intentionally blank...
Page 30: ...Serinus 44 User Manual 3 0 Page 30 This page is intentionally blank...
Page 64: ...Serinus 44 User Manual 3 0 Page 64 This page is intentionally blank...
Page 112: ...Serinus 44 User Manual 3 0 Page 112 This page is intentionally blank...
Page 128: ...Serinus 44 User Manual 3 0 Page 128 This page is intentionally blank...
Page 137: ...Troubleshooting Page 137 7 6 USB Memory Stick Failure Figure 71 USB Memory Stick Failure...
Page 140: ...Serinus 44 User Manual 3 0 Page 140 This page is intentionally blank...
Page 146: ...Serinus 44 User Manual 3 0 Page 146 This page is intentionally blank...
Page 150: ...Serinus 44 User Manual 3 0 Page 150 9 5 Plumbing Schematic D020018...
Page 151: ...Parts List and Schematics Page 151 9 6 Block Wiring Schematic D020104...
Page 152: ...Serinus 44 User Manual 3 0 Page 152 9 7 Ozone Generator Assembly H011107...
Page 153: ...Parts List and Schematics Page 153 9 8 Optical Bench Assembly H011200...
Page 154: ...Serinus 44 User Manual 3 0 Page 154 9 9 Manifold Block Assembly H011302...
Page 155: ...Parts List and Schematics Page 155 9 10 Calibration Manifold Assembly H010013 01...
Page 156: ...Serinus 44 User Manual 3 0 Page 156 9 11 Valve Assembly H010042...
Page 173: ...Parts List and Schematics Page 173 This page is intentionally blank...