Serinus Cal 3000
N/A
O
3
N/A
4.4.1.1
Analog Outputs Voltage Calibration
Equipment Required
Multimeter (set to volts)
Male 25 pin connector with cable
Procedure
1.
Open -
Main Menu
Communications Menu
Analog Output Menu
(refer to Section 3.4.24).
2.
Select -
Output Mode
Voltage
.
3.
Connect a multimeter (using an appropriate adaptor or probes on the multimeter) to the ground
(pin 24) and the relevant output pin (pin 10).
4.
Edit -
0.5V Calibration
- (until the multimeter reads 0.500 ±0.002) - Accept.
5.
Edit -
5.0V Calibration
- (until the multimeter reads 5.00 ±0.002) - Accept.
4.4.1.2
Analog Outputs Current Calibration
Equipment Required
Multimeter (set to mA)
Male 25 pin connector with cable
Procedure
1.
Open -
Main Menu
Communications Menu
Analog Output Menu
(refer to Section 3.4.24).
2.
Select -
Output Mode
Current
.
3.
Connect a multimeter (using an appropriate adaptor or probes on the multimeter) to the ground
(pin 24) and the relevant output pin (pin 10).
4.
Edit -
4mA Calibration
- (until the multimeter reads 4mA ±0.01) - Accept.
5.
Edit -
20mA Calibration
- (until the multimeter reads 20mA ±0.01) - Accept.
4.4.2
Analog Inputs
The instrument is also equipped with three analog inputs with resolution of 15 bits plus polarity,
accepting a voltage between 0-5V. These go directly to the microprocessor and should be protected
to ensure static/high voltage does damage the main controller PCA (instrument warranty does not
cover damage from external inputs).
Summary of Contents for Serinus 44
Page 1: ...Serinus 44 Ammonia Oxides of Nitrogen Analyser User Manual Version 3 0 www ecotech com...
Page 2: ...Serinus 44 User Manual 3 0 Page 2 This page is intentionally blank...
Page 30: ...Serinus 44 User Manual 3 0 Page 30 This page is intentionally blank...
Page 64: ...Serinus 44 User Manual 3 0 Page 64 This page is intentionally blank...
Page 112: ...Serinus 44 User Manual 3 0 Page 112 This page is intentionally blank...
Page 128: ...Serinus 44 User Manual 3 0 Page 128 This page is intentionally blank...
Page 137: ...Troubleshooting Page 137 7 6 USB Memory Stick Failure Figure 71 USB Memory Stick Failure...
Page 140: ...Serinus 44 User Manual 3 0 Page 140 This page is intentionally blank...
Page 146: ...Serinus 44 User Manual 3 0 Page 146 This page is intentionally blank...
Page 150: ...Serinus 44 User Manual 3 0 Page 150 9 5 Plumbing Schematic D020018...
Page 151: ...Parts List and Schematics Page 151 9 6 Block Wiring Schematic D020104...
Page 152: ...Serinus 44 User Manual 3 0 Page 152 9 7 Ozone Generator Assembly H011107...
Page 153: ...Parts List and Schematics Page 153 9 8 Optical Bench Assembly H011200...
Page 154: ...Serinus 44 User Manual 3 0 Page 154 9 9 Manifold Block Assembly H011302...
Page 155: ...Parts List and Schematics Page 155 9 10 Calibration Manifold Assembly H010013 01...
Page 156: ...Serinus 44 User Manual 3 0 Page 156 9 11 Valve Assembly H010042...
Page 173: ...Parts List and Schematics Page 173 This page is intentionally blank...