FloormapX System Overview
21
Scanning head
The scanning head connects to the mainframe and consists primarily of the variable strength
magnetic bridge, the MFL and STARS sensor head modules. The main components of the scanning
head are identified below. The scanning head product and serial number label is located at the rear
of the unit and is only visible when the mainframe is not attached.
Figure 4: Scanning head overview
Figure 5: Scanning head product and serial number label
Magnet Position Indicators
On the right-hand side of the scanning bridge, when viewing the system from the front, is the
magnetic position indicator. The indicator identifies how much magnetic strength is imparted into
the inspection surface between a range of 0 to 100. When the indicator points to the zero ‘0’ position
then the FloormapX magnets are deemed off and do not impart magnetism into the inspection
surface. When the magnetic indicator points at any other value then the FloormapX magnets will
impart magnetism into the inspection surface or any other ferrous surface or object. The greater the
value on the magnetic position indicator the stronger the magnetic attraction.
NOTE
When at the zero position there will still be some low-level residual magnetism around the
system and there are places where magnetically attract small ferromagnetic items.
Scanning Bridge Height Adjuster
The scanning bridge can be set at two different height positions, low or high, using the height
adjustment level.
Before setting the bridge height either set the magnets to the zero position (magnets off) or lie the
system in a horizontal position so the scanning bridge is clear of the inspection surface.
Summary of Contents for Floormap X
Page 1: ...User s Manual Floormap X MFL Array Tank Floor Inspection Solution...
Page 10: ...10 Chapter 1 Precautions and conventions...
Page 16: ...General Precautions and Conventions 16 1 5 EU Declaration of Conformity...
Page 19: ...19 Chapter 2 FloormapX System Overview...
Page 35: ...35 Chapter 3 Inspection Preparation Setup and Operation...
Page 43: ...43 Chapter 4 SIMS GO Data Acquisition Software...
Page 69: ...69 Chapter 5 SIMS PRO...