SIMS PRO
97
Plate Modification
The plate modification options available is determined by what type of plate is selected. When a plate
setup change has been made the plate become hatched red to identify a modification has been
made.
IMPORTANT
Verify the location of all scans and indications after setup changes are applied.
NOTE
Plate setup changes only update the planned scan positions. The position of any curved or
added scan, or manually altered planned scans do not get update.
Plate Orientation
The plate orientation on the modification ribbon can be changed between either a horizontal or a
vertical plate.
Plate Reference
The plate reference corner on the modification ribbon can be changed between either the Top Left,
Top Right, Bottom Left or Bottom Right corner.
Scan Pattern
The scan pattern on the modification ribbon can be changed between either a Parallel or a Raster
scanning format.
Plate Size
The dimensions entered for the plate during data acquisition software can be changed in the plate
properties.
NOTE
Scans do not move when the plate size is changed. If scans exist outside of the plate, it is not
possible to exit plate view.
Scan Position
A scan can be manually repositioned on the plate. With Enable Plate Modification ticked, select a
scan then click, hold, and move to the desired location. When a scan position change has been made
the plate become hatched red to identify a modification has been made.
NOTE
The start of the scan is indicated by the red corner. The arc of a curved scan cannot be
altered.
Reset Plate Modification
Click
Reset Plate Modification
on the modifications ribbon to revert the plate setup and size back to
its initial setup options and clears the red hatching of the plate.
NOTE
Reset plate modification does not return curved or added scans back to their initial position.
Summary of Contents for Floormap X
Page 1: ...User s Manual Floormap X MFL Array Tank Floor Inspection Solution...
Page 10: ...10 Chapter 1 Precautions and conventions...
Page 16: ...General Precautions and Conventions 16 1 5 EU Declaration of Conformity...
Page 19: ...19 Chapter 2 FloormapX System Overview...
Page 35: ...35 Chapter 3 Inspection Preparation Setup and Operation...
Page 43: ...43 Chapter 4 SIMS GO Data Acquisition Software...
Page 69: ...69 Chapter 5 SIMS PRO...