25
6.
LYNCS-CM Overview
Top View
Figure 17: LYNCS-CM frame top overview
1.
On board controls
This button is used for either the
Increment / Encoder Reset
functionality
or
Pause / Resume
functionality.
To use the
Pause / Resume
functionality,
plug the encoder umbilical cable directly
into the instrument.
To use the
Increment / Encoder Reset
functionality, connect the supplied
encoder adapter cable between the
encoder umbilical and instrument.
*Adapter supplied with LYNCS-FULL, LYNCS-CM.
2.
Scanner rail holes
Allows the corrosion mapping mount to
be locked along the scanning frame by
clicking into each hole position. The
spacing between the holes in the metric
bars (667mm) is set at 50mm and for
the imperial bars (26.26in) the spacing is
set at 2in.
3.
Water box
Designed to house the immersion phased
array probes with external dimensions (L
x W x H) of 73 x 24 x 25 mm (2.87 x 0.94
x 0.98 in). Water hose connections and
pressure release valves are fitted to
maintain a water column. The water box
wedge column of either 15mm or 30mm.
can be selected depending on the
thickness of the test part and
application.
Isometric View
Figure 18: LYNCS-CM scanner isometric overview
4.
Corrosion mapping mount
Used to house the supplied water box.
The corrosion mapping mount fits
between the scanning frame and slides
along it. Its position can be locked
manually using the cam lever lock or via
the spring-loaded lever lock which clicks
into each hole in along the length of the
scanning rail. The lever lock must sit on
the same side as the rail with the drilled
holes.
5.
Water box clamping mechanism
Used to secure the water box at the
desired position in the corrosion mapping
mount.
Tighten the thumbscrew
clockwise to clamp the water box.
Loosen counter clockwise to release.