PPS20
–
User Manual Rev. 1.10
EDFelectronics JRMM Sp. z o.o. sp.k.
Rybnicka 64, Radlin 44-310, Poland
; e-mail: info@edfelectronics.eu tel.: +48 604 343 504
34
field into the blinking ( about 1/s ). Now output power set is possible
–
use rotary
encoder or arrow buttons to change it.
Fig. 4.2.5
NOTE:
Type of operation (Constant Power, Voltage or Current) can be selected
only when high voltage HV is not active. When HV stays on, then value of selected
mode can be set only.
NOTE:
Maximum adjustable output
power value is limited by “Limit Settings” –
refer to chapter 4.3.4 for details.
NOTE:
Input data for Constant Power, Voltage or Current operation are set in the
same way for both of mode of operation: DC or PulseDC
4.2.4
P
OWER
C
ALCULATIONS
Output power of PPS20 is different for each mode operation
–
DC or PulseDC . It
is really important to understand how it works.
For DC mode operation it is really easy, and output power is calculated according
to the following formula:
P
out
=
U
out
x
I
out
where:
P
out
–
nominal output power [W]
U
out
–
nominal output voltage [V],
I
out
–
nominal output current [A],
Maximum output voltage for DC operation is 1000 V (1200V during ignition), and
max output current is 3A, however max output power is 1500W, and PPS20 is
operating according to the following characteristic: