A
AC
C V
V
1. Frequency range for ACV: 40Hz~400Hz.
2. Response for ACV: Average, calibrated in rms of sine wave.
- Overload protection: 1000V DC or 700V AC rms.
- Input impedance (nominal):
DC voltage: >10M
Ω
AC voltage: >10M
Ω
- Common mode rejection ratio:
DC voltage: >100dB at DC, 50 or 60 Hz;
AC voltage: >60dB at DC, 50 or 60 Hz
- Normal mode rejection ratio: DC voltage: >45dB at 50 or 60Hz
FREQUENCY (10Hz-1MHz)
• Overload protection: 250V DC or 250V AC rms
RESISTANCE
• Overload protection: 250V DC or 250V AC rms
RANGE
RESOLUTION
ACCURACY
99.99Hz
0.01Hz
999.9Hz 0.1Hz
9.999Hz
0.001kHz
± (0.1% of rdg + 3 digits)
99.99kHz
0.01kHz
999.9kHz
0.1kHz
RANGE
RESOLUTION
ACCURACY
600.0
Ω
0.1
Ω
± (0.5% of rdg + 3 digits)
6.000k
Ω
1
Ω
60.00k
Ω
10
Ω
± (0.5% of rdg + 2 digits)
600.0k
Ω
100
Ω
6.000M
Ω
1k
Ω
60.00M
Ω
10k
Ω
± (1.5% of rdg + 3 digits)
RANGE
RESOLUTION
ACCURACY
600mV
0.1mV
± (3.0% of rdg + 3 digits)
Manual Range only
6V 1mV
60V
10mV
± (1.0% of rdg + 3 digits)
600V
100mV
700V
1V
±(1.5% of rdg + 3 digits)