background image

2

2. Indications for Use

Pericardial valves are indicated for use in patients suffering
from valvular heart disease. Aortic valvular heart disease is a
condition involving any of the following: obstruction of the
aortic heart valve or stenosis; leakage of the aortic valve,
known as regurgitation, incompetence, or insufficiency; and
combinations of the two, sometimes referred to as mixed
disease or combined lesions.

Aortic valvular heart disease may be caused by any number of
factors, including congenital abnormalities, infection by various
microorganisms, degenerative calcification, and rheumatic
heart disease.

Pericardial valves are used particularly in those patients for
whom long-term anticoagulation is contraindicated or who may
be difficult to maintain on anticoagulation therapy.

The Carpentier-Edwards PERIMOUNT Magna pericardial
bioprosthesis is intended for use in patients whose aortic
valvular disease is sufficiently advanced to warrant replacement
of their natural valve with a prosthetic one. It is also intended
for use in patients with a previously implanted aortic valve
prosthesis that is no longer functioning adequately and
requires replacement. In the latter case, the previously
implanted prosthesis is surgically excised and replaced by the
replacement prosthesis. The valve can be implanted in either
the supra-annular or intra-annular position.

3. Contraindications

Do not use if surgeon believes such would be contrary to the
best interests of the patient. The actual decision for or against
the use of this valve must remain with the surgeon who can
evaluate all the various risks involved, including the anatomy
and pathology observed at the time of surgery.

4. Warnings

For Single Use Only

DO NOT RESTERILIZE THE VALVE BY ANY METHOD. 

Exposure

of the bioprosthesis or container to irradiation, steam, ethylene
oxide, or other chemical sterilants will render the bioprosthesis
unfit for use.

DO NOT FREEZE OR EXPOSE THE VALVE TO EXTREME HEAT.

Each bioprosthesis in its jar is shipped in a molded foam
enclosure containing a temperature indicator, which is intended
for monitoring the temperature that the device is exposed to
during transit. If the indicator has been activated, indicating the
valve has been exposed to freezing temperatures or has had
prolonged exposure to heat, do not use the valve. Please refer
to the Storage section (10.3) for further instructions.

DO NOT USE

the bioprosthesis if the tamper evident seal is broken.

DO NOT USE

if expiration date has elapsed.

DO NOT USE

the bioprosthesis if the container is leaking,

damaged, or the glutaraldehyde solution does not completely
cover the bioprosthesis.

DO NOT EXPOSE

the valve to any solutions, chemicals,

antibiotics, etc. except for the storage solution or sterile
physiological saline solution, as irreparable damage to the leaflet
tissue may result that is not apparent under visual inspection.

DO NOT ALLOW

the valve tissue to dry. It must be kept moist

at all times. Maintain tissue moisture with sterile physiological
saline irrigation on both sides of the leaflet tissue.

DO NOT PASS CATHETERS

, transvenous pacing leads, or any

surgical instrument across the valve since it may cause tissue
damage.

DO NOT USE

the valve if it has been dropped, damaged, or

mishandled in any way. Should a bioprosthesis be damaged
during insertion, do not attempt repair.

DO NOT HANDLE

the leaflet tissue of the bioprosthesis with

instruments or cause any damage to the valve tissue. Even the
most minor tissue perforation may enlarge in time to produce
significant impairment of valve function.

Clinical data which establishes the safety and efficacy of the
valve for use in patients under the age of 20 is not available;
therefore, we recommend careful consideration of its use in
younger patients.

The decision to use a tissue valve must ultimately be made by
the physician on an individual basis after a careful evaluation of
the short- and long-term risks and benefits to the patient and
consideration of alternative methods of treatment. Long-term
durability has not been established for bioprostheses.

Serious adverse events, sometimes leading to replacement of
the valve and/or death, may be associated with the use of
prosthetic valves (see 

6. Adverse Events

). A full explanation of

the benefits and risks should be given to each prospective
patient before surgery.

Note:

Bioprostheses should be used with caution in the

presence of severe systemic hypertension or when the
anticipated patient longevity is longer than the known longevity
of the prosthesis (see 

7. Clinical Studies

).

Careful and continuous medical follow-up (at least by an annual
visit to the physician) is advised so that valve-related
complications, particularly those related to material failure, can
be diagnosed and properly managed.

Recipients of prosthetic heart valves who are undergoing dental
procedures should receive prophylactic antibiotic therapy to
minimize the possibility of prosthetic infection.

Bioprosthetic heart valve recipients should be maintained on
anticoagulant therapy (except where contraindicated) during the
initial healing stages after implantation, approximately 2 to
3 months. Anticoagulants should then be discontinued over a
period of 10 days, except in those patients for whom indefinite
anticoagulant protection is indicated, i.e., in the absence of sinus
rhythm and in patients with a dilated left atrium, calcification of
the atrial wall, or history of previous atrial thrombus. However,
the appropriate anticoagulation therapy must be determined by
the physician on an individual basis (Ref. 1).

149256001A_EN_ES_PT  16.2.2005  14:30 Uhr  Seite 2

Summary of Contents for 3000TFX PERIMOUNT Magna Aortic

Page 1: ...logy and mechanics of natural heart valves and reported experience with implantation of unstented homografts Refs 5 7 The lightweight wireform frame is made of Elgiloy a corrosion resistant alloy chos...

Page 2: ...ons chemicals antibiotics etc except for the storage solution or sterile physiological saline solution as irreparable damage to the leaflet tissue may result that is not apparent under visual inspecti...

Page 3: ...ression of the valve orifice When using interrupted sutures it is important to cut the sutures close to the knots and to ensure that exposed suture tails will not come into contact with the leaflet ti...

Page 4: ...available on 719 patients requiring isolated aortic valve replacement AVR with the Model 2700 Carpentier Edwards pericardial bioprosthesis with mean follow up of 3 9 years indicate overall actuarial...

Page 5: ...HA functional classification has also been demonstrated postoperatively Forty five percent of the patients are in NYHA Functional Class I at 12 years post implant with the Carpentier Edwards pericardi...

Page 6: ...should contain enough buffered glutaraldehyde storage solution to cover the prosthesis The contents of the jar should be handled in an aseptic manner to prevent contamination Using gloved hand attach...

Page 7: ...recautions Caution Because of the intense temperature and lighting conditions in the operating field the bioprosthesis should be irrigated frequently every 1 to 2 minutes is recommended on both sides...

Page 8: ...ion is not necessary under these circumstances 12 Patient Information 12 1 Registration Information An Implantation Data Card is included in each device package for patient registration After implanta...

Page 9: ...nte en pacientes con ra ces a rticas peque as se ha reducido el di metro de la biopr tesis peric rdica PERIMOUNT Magna de Carpentier Edwards 2 Indicaciones de uso Las v lvulas peric rdicas est n indic...

Page 10: ...entre 2 y 3 meses Transcurrido este per odo se debe suspender el uso de anticoagulantes durante un per odo de 10 d as excepto en aquellos pacientes para los que est indicada una protecci n anticoagula...

Page 11: ...s que contengan calcio en el postoperatorio as como la ingesti n excesiva de leche y derivados l cteos en los ni os Los estudios realizados en animales ref 11 demuestran que la concentraci n elevada d...

Page 12: ...alg n familiar o el m dico de su zona En la tabla 1 se resumen los porcentajes de complicaciones operatorias y postoperatorias en la poblaci n sometida a AVR aislada y a DVR Los porcentajes de complic...

Page 13: ...ds Lifesciences LLC One Edwards Way Irvine CA 92614 5686 EE UU 8 Individualizaci n del tratamiento Los receptores de v lvulas card acas bioprot sicas deber n someterse a terapia de anticoagulaci n exc...

Page 14: ...por s sola del 100 contra todos los posibles contaminantes No se debe intentar volver a esterilizar la biopr tesis peric rdica PERIMOUNT Magna de Carpentier Edwards Precauci n Tanto la v lvula como l...

Page 15: ...i n intra anular se puede utilizar el extremo cil ndrico o de la r plica de la v lvula del calibrador modelo 1130 Para realizar la calibraci n correctamente el calibrador debe estar paralelo al plano...

Page 16: ...olver una v lvula recuperada p ngase en contacto con su especialista local en v lvulas Las v lvulas explantadas deber n colocarse en un fijador histol gico adecuado como formol al 10 o glutaraldeh do...

Page 17: ...tier Edwards foi reduzido para facilitar a implanta o em doentes com ra zes a rticas pequenas 2 Indica es de Utiliza o As v lvulas peric rdicas est o indicadas para utiliza o em doentes que sofram de...

Page 18: ...vulares card acas devem receber uma profilaxia antibi tica adequada durante tratamentos odontol gicos a fim de reduzir o risco de infec o da pr tese Os portadores de pr teses biol gicas valvulares car...

Page 19: ...r ser devolvida sem ser utilizada Esta etiqueta n o dever ser separada da v lvula at imediatamente antes da implanta o Dever se ter cuidado para evitar cortar ou rasgar o tecido do anel de sutura dura...

Page 20: ...tando de dupla substitui o da v lvula DSV com um acompanhamento m dio de 3 7 anos indicam uma taxa actuarial global de sobreviv ncia a 6 anos de 67 2 6 5 Estes dados do grupo de doentes pr aprova o fo...

Page 21: ...doentes que necessitaram de nova opera o explante e a disfun o valvular foi a causa de morte em dois doentes Embora a taxa de sobreviv ncia global dos doentes seja de 45 a 12 anos a taxa de aus ncia d...

Page 22: ...o por parte do cliente ser o considerados como fruto da exposi o a condi es ambientais da responsabilidade do cliente e como tal ser o sujeitos a substitui o a cargo do cliente 10 3 Conserva o As pr t...

Page 23: ...as que o cirurgi o considere necess rias 2 Retirar cirurgicamente as calcifica es do annulus a fim de obter um alojamento adequado do anel de sutura 3 Medir o tamanho do annulus utilizando exclusivame...

Page 24: ...mente a geometria do anel de sutura da v lvula Figura 4b No outro lado do punho existe uma extremidade de r plica da v lvula que reflecte a geometria do anel de sutura da v lvula assim como a altura e...

Page 25: ...am se a utiliza o para os registos hospitalares e do cirurgi o Quando o Registo de Implantes em Doentes receber o cart o ser emitido um cart o de identifica o tamanho carteira que ser enviado ao doent...

Page 26: ...ke Medical Books New York 1982 pp 25 34 17 Ferrans V J et al Structural Changes in Glutaraldehyde Treated Porcine Heterografts Used as Substitute Cardiac Valves Am J Cardiol 41 1159 1184 1978 18 Forfa...

Page 27: ...cation of Tissue Heart Valve Prostheses J Thorac Cardiovasc Surg 62 5 683 689 and 693 695 1971 48 Relland J et al The Third Generation Carpentier Edwards Bioprosthesis Early Results JACC 6 5 1149 1154...

Page 28: ...5 mm 27 mm 29 mm A Di metro del stent malla 19 21 23 25 27 29 B Di metro interior D I del stent 18 20 22 24 26 28 C Altura del perfil 14 15 16 17 18 19 D Di metro del anillo de sutura externo 24 26 28...

Page 29: ...5 1 0 0 0 8 NA Segunda intervenci n relacionada con la v lvula 0 7 0 1 99 8 0 4 0 0 NA Todas las segundas intervenciones 22 4 1 8 75 4 1 8 34 3 2 3 NA Tromboembolismo relacionado con la v lvula 3 1 1...

Page 30: ...9 2 Velocidad M seg media D E 2 80 0 49 2 56 0 46 2 36 0 42 2 15 0 56 2 09 0 27 2 08 0 1 2 46 0 50 n 12 21 15 7 3 3 61 rango 1 90 3 60 1 90 3 90 1 39 2 86 1 00 2 60 1 90 2 40 2 05 2 10 1 00 3 90 Pico...

Page 31: ...acia grados funcionales de la NYHA modelo 2700 Preoperatorio Postoperatorio Clase funcional NYHA Clase funcional NYHA No I II III IV Expiraci n disponible I 18 19 9 II 140 37 35 15 III 181 48 4 1 72 2...

Page 32: ...icaci n N de incidencias N de incidencias Tromboembolismo Trombo 8 3 0 31 1 45 Endocarditis 0 0 7 0 33 Disfunci n de la v lvula 1 0 37 34 1 60 Fuga perivalvular 1 0 37 4 0 19 Complicaci n por anticoag...

Page 33: ...ra 3 A rtico CP1036 45 Figure 4a Aortic Sizer Figura 4a Calibrador a rtico Figura 4a Medidor A rtico CP1036 33 Figure 1 Aortic Figura 1 A rtica Figura 1 A rtico CP1036 46 CP1036 48 CP1036 47 Figure 2...

Page 34: ...Medi o Intra Anular CP1036 39 CP1036 40 Figure 6b Intra Annular Measurement Figura 6b Calibraci n intra anular Figura 6b Medi o Intra Anular CP1036 32 Figure 6c Intra Annular Placement Figura 6c Coloc...

Page 35: ...vapor o calor vapore o a door stoom of eller t rsteriliseret med nga eller por Vapor Dry Heat la chaleur s che trockener W rme seco calore secco droge warmte torrv rme ou Calor Seco Serial Number Num...

Page 36: ...es LLC One Edwards Way Irvine CA 92614 5686 USA 01 05 149256001 Rev A Copyright 2005 Edwards Lifesciences LLC All rights reserved 149256001A Edwards Lifesciences Services GmbH Edisonstr 6 D 85716 Unte...

Reviews: