OPERATION
4.2
GasCheck G3 running screen
The GasCheck G3 displays the 'running screen' whenever the instrument is used to detect leaking gas.
describes the information and various icons displayed.
Figure 5 - Running screen
4.3
Using the GasCheck G3
CAUTION
Ensure that the ambient air is clean and free from any tracer gas when the GasCheck G3 is switched on. Failure
to do so will result in inaccurate zero readings.
CAUTION
Do not touch the probe or brass sensor housing. Any changes in heat can result in significant changes in signal
causing false readings.
CAUTION
Avoid placing the instrument on wet or dirty surfaces. Blocking of the probe will result in instrument failure.
CAUTION
Do not bend the probes. The accuracy of the GasCheck G3 will be affected.
1. Gas selection
2. Sensitivity
3. Sound indicator
4. Battery status indicator
5. Backlight indicator
6. Peak hold indicator
7. Data symbol indication of leak rate stored in memory
8. Bar graph indication of leak rate
9. Leak rate