D397-21-880 Issue G
Page 24
© Edwards Limited 2009. All rights reserved.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Operation
Figure 19 - Gauge setup screen APGX-M
APGX has an in-built error monitoring capability. (Refer to
Note:
There is a
+/−
50 mV tolerance on the gauge outputs shown in
.
4.8.5
Active Thermocouple Gauge (ATC-E) control
Note:
For a detailed specification and instructions regarding the use of the ATC-E, refer to the Instruction Manual
(D351-08-880).
Active Thermocouple Gauges are low cost, gas dependent, high pressure, thermal conductivity gauges. The ATC-E
electronics module is compatible with two types of thermocouple gauge tube to give a wide measuring range. (Refer
to
).
Before connecting to TIC, the ATC-E must be configured for the type of gauge tube (ATC-D or ATC-M) to be used. This
is done by selecting the body colour of the tube with the two-position switch.
Table 12 - Error/diagnostic monitoring APGX
Diagnostic messages
Description
Over Range
(Gauge output
=
9.750 V)
Indicates that there is a fault with the gauge and it should be replaced.
Under Range
(Gauge output
<
0.300 V)
Indicates that there is a fault with the gauge and it should be replaced.
Filament Fail
(Gauge output
=
9.500 V)
Indicates that the gauge filament is broken.
The gauge should be replaced.
Cal Error
(Gauge output
=
9.600 V)
Indicates that the gauge has gone out of calibration.
This could be because of the wrong gas type being selected, pressing the
atmosphere and vacuum adjustment button in mid-range or contamination of the
gauge.
The gauge should be re-adjusted as described in
If the fault persists the gauge should be replaced.
No Tube
(Gauge output
=
9.700 V)
Indicates that the replaceable gauge tube is missing or incorrectly fitted. Check
the tube and ensure that the two fixing screws are properly secured.
Note:
APGX-H only.
Table 13 - ATC-D and ATC-M gauge tubes
Tube name
Tube part number
Pressure measuring range
ATC-D
D355-12-000
65 to 6.5 x 10
-2
mbar (50 to 5 x 10
-2
Torr)
ATC-M
D355-13-000
1.3 to 1.3 x 10
-3
mbar (1 to 1 x 10
-3