© Edwards Limited 2009. All rights reserved.
Page 3
Edwards and the Edwards logo are trademarks of Edwards Limited.
Technical data
D397-21-880 Issue G
2
Technical data
2.1
Electrical data
2.2
Operating and storage data
2.3
Mechanical data
2.4
Connections
2.4.1
Active gauge connectors
Connector type
CEE/IEC 320
Electrical supply
90 to 264 V a.c. 47 to 63 Hz
Power consumption
TIC Instrument Controller
55 VA maximum (D397-00-000)
Peak inrush current
5 A at 110 V a.c
D397-00-000
12 A at 240 V a.c
D397-00-000
Fuse
The unit is self-protecting and has no user replaceable
fuse. The unit will recover once any overload is removed.
Earth Stud
M4
Ambient operating temperature range
0 °C to 40 °C
Ambient storage temperature range
-30 °C to 70 °C
Maximum ambient operating humidity
Max 90% RH non condensing at 40 °C
Maximum operating altitude
3000 m max
IP rating
20
Weight
TIC Instrument Controller
1.3 kg
Connector type
FCC/RJ45, 8-way (refer to
)
Power supply
24 V d.c.
Maximum power rating
21 W total for the three gauges
Input voltage range
-0.5 V to 15 V
Output ID current
33
µ
A, 0 V to 13 V
Control output
active:
<
0.8 V d.c. (2 mA max)
inactive:
open (internal pull-up to 24 V)
Control input
low:
<
2.0 V d.c. (I
out
<
160
µ
A)
high:
>
3.5 V d.c. (internally pulled up to 24 V)