Pneumatically operated breastbone saw EFA 57
Operating instructions
EN
Part no. 110 885 700 – Rev. 00 – 01.2020
17 of 34
3
Connecting the EFA pneumatically operated breastbone saw
Attention!
Device damage!
Operate the EFA pneumatically operated breastbone saw only with clean, water-free and oiled
air!
Insert the air hose (1) as far as it will go into the screw-in fitting (2) on the valve housing (3) of the EFA
pneumatically operated breastbone saw.
Screw the air-outlet hose (4) onto the valve housing (3) of the EFA pneumatically operated breastbone
saw.
Connect the EFA pneumatically operated breastbone saw to the oil mist lubricator (6) with the thread
nipple of the air hose (5) (with filter pressure reducer).
Make sure that the oil mist lubricator is assembled in the following sequence: water separator (7), pressure
gauge (8), oil mist unit (9).
The oil mist lubricator and the air hose are not included in the scope of supply but can be
purchased on request from our Sales department at
Schmid & Wezel GmbH
.
Air hose:
Part no. 311 366 811
Oil mist lubricator: Part no. 311 367 043
To ensure lubrication of the pneumatic motor, the oil mist lubricator must be set to 2 – 3 drops of oil per minute
(or one drop of oil for approx. 15 cuts).
If the compressed air connection is faulty and there is a high water content in the compressed air, water can
escape at the connection point.
Therefore, make sure that the hose connection is correct and that compressed air with a low water content is
used (the water separator must be emptied regularly).
If no dried air is available, a water absorber must be installed by the customer.