56
- Take out the battery when it is not using for a long time.
- Do not use or store the Meter in a place of humidity, high temperature, explosive, inflammable
and strong magnetic field.
B. Replacing the Fuses
Warning!
To avoid electrical shock or arc blast, or personal injury or damage to the Meter, use
specified fuses ONLY in accordance with the following procedure
Follow Figure 6-1 and proceed as follows to replace the Meter’s fuse:
- Turn the rotary switch to OFF and remove all connections from the terminals.
- Remove the 5 screws from the case bottom.
- Remove the fuse by gently prying one end loose, then take out the fuse from its bracket.
- Install ONLY replacement fuses with the identical type and specification as follows and make
sure the fuse is fixed firmly in the bracket.
Fuse 1: 0.5A, 250V, fast type fuse, Ø5×20mm
Fuse 2: 10A, 250V, fast type fuse, Ø5×20mm
- Rejoin the case bottom and case top, and install the 5 screws.
Replacement of the fuses is seldom required. Burning of a fuse always results from improper
operation.
Figure 6-1. Fuse Replacement
C. Replacing the Battery
Warning!
To avoid false readings, which could lead to possible
electric shock or personal injury, replace the battery
as soon as the battery indicator “
” appears.