59
STORE Readings
Up to 100 readings for COD. 51254 may be saved by
the user in a memory. These readings may be viewed by
using Recall feature.
Detailed Accuracy Specifications
Accuracy: ([% of reading] + [number of least significant digits), guarantee for 1 year.
Operating temperature: 18ºC~28ºC
Relative humidity: 75%RH
A. DC Voltage
Range
Resolution
Accuracy
Overload
Protection
Input
Impedance
COD. 51253
COD. 51254
200mV
0.01mV
±(0.1%+8)
±(0.05%+5)
1000V
Around 2.5G
2V
0.0001V
±(0.08%+5)
Around 10M
20V
0.001V
200V
0.01V
1000V
0.1V
±(0.15%+8)
±(0.1%+8)
B. AC Voltage (AC+DC measurement is available)
Range
Resolution
Bandwidth
Accuracy
Input
Impedance
COD. 51253
COD. 51254
2V
0.0001V
45Hz~1kHz
±(0.8%+40)
±(0.6%+40)
Approx 10M
1kHz~10kHz
±(1.8%+40)
±(1.6%+40)
10kHz~100kHz
±(7%+40)
±(7%+40)
20V
0.001V
45Hz~1kHz
±(0.8%+40)
±(0.6%+40)
1kHz~10kHz
±(2%+40)
±(1.8%+40)
10kHz~100kHz
±(7%+40)
±(7%+40)
200V
0.01V
45Hz~1kHz
±(0.8%+40)
±(0.6%+40)
1kHz~10kHz
±(5%+40)
±(5%+40)
10kHz~100kHz
Not Specified
Not Specified
1000V
0.1V
45Hz~1kHz
±(1.5%+40)
±(1.2%+40)
1kHz~5kHz
±(6%+40)
5kHz~10kHz
±(10%+40)
Remarks:
- Overload Protection: 1000V.
- Display:
a) True rms are valid from 10% of range to 100% of range
b) AC crest factor can be up to 3.0 except 1000V where it is 1.5.
c) A residual reading of 80 digits with test leads shorted, will not affect stated accuracy.
d) When frequency is lower than 100kHz, the accuracy guarantee range 10%-100%.
e) When making AC+DC measurment, the accuray need to add (1%+ 35 digits) of reading