61760V04-2012-09
Seite / Page 25 von / of 28
The unit is on – no function
The unit’s fuse is interfered. Possible reason: Excess voltage from the power line or an error inside the
unit.
Solution: Check the unit’s fuse at the back and replace it, if necessary. Only use a fuse with the
stated value and the stated response properties because, otherwise, a dangerous situation might
occur. If the fuse activates, again, or if there is still no function, please send the unit to the after-sale
service.
The internal irreversible over-temperature protection has not actuated
Solution: Send the unit to the after-sale service.
The unit cannot be started with the button “T1” and is sending a signal tone
The cover is not closed.
Solution: Close the cover. If the error does not disappear the cover switch is defect and must be
replaced by the after-sale service.
The unit cannot be operated by pressing the button “START CLOCK”
The cover is not closed.
Solution: Close the cover. If the error has not disappeared, then the cover switch is defect and must
be replaced by the after-sale service.
“E” is displayed
If “E” is displayed, an internal error has occurred. The over-temperature
protection might have been actuated. This can happen, when the unit
has been operated without any filling or with an insufficient filling.
Solution: Turn the unit off, allow it to cool down and fill it sufficiently.
Another possible reason could be a defect on one of the sensors.
Solution: Send the unit to the after-sale service.
The error also occurs when the unit is operated at a room temperatures below + 10 °C (50 °F).
Solution: Operate the unit at higher temperatures. Some time might pass until the unit has adapted
to the higher temperature.
The agitator does not start immediately after running the program
This is not an error. The unit, first, heats up until the ground sensor reports a standard temperature.
Only then is the agitator released. This function is necessary to detach solidified material, if any.
Solution: Wait until a standard temperature is reached for the first time.