position, as the result, the model is not completely
dried.It is recommended to use longer duration of
drying period for complicated models. Also, please face
up the surface of the model that is not easy to dry as
well as to tilt the model appropriately to avoid storing a
large amount of cleaning fluid in the groove of the
model.
2.
The water level button flashes and the buzzer
alarms
① If the cleaning fluid is insufficient, just replenish the
cleaning fluid and restart the device.
②if the water inlet tube above the liquid surface, pleaes
insert it to the bottom of the liquid and restart the
device.
3. All button lights flash and buzzer alarms
High level sensor detects signal. Power off the unit, wipe
the position of the high level sensor, clean and restart
4.
Drain pump working after press start button.
①
When there are liquids inside, the device will
discharge first, and then proceed to the normal
operation process.
② There are too many water drops inside of the the
device (especially low liquid level), just wipe and clean
the internal guts wall.