5. SYSTEM INSTALLATION AND CHECK
Warning!
• The waterproof grade of the equipment is IPX4 (no adverse effect on splashing water in all directions). Splash
any liquid on the equipment may damage it. If you accidentally splash liquid on the equipment, please
immediately turn off the power and contact your service representative.
• To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective
earth.
Warning!
1. All instrument plugs of this system shall be connected into the power socket with protectively earth on the
wall and the socket must meet the requirement of power rating of instrument. Use of multiple portable socket-
outlets may affect protective earth to make leakage currents exceed the safety requirements.
2. Please follow the correct electrical connection method to connect the power supply and earth, otherwise there
will be danger of electric shock. Do not connect the grounding wire to any gas pipe or water pipe, or it may
cause bad grounding and danger of explosion.
3. Prohibit the live parts of the equipment or other devices (such as various signal input and output ports, etc.)
contact with the patient, if this equipment or other equipment has failure, the patient will have danger of
electric shock.
4. Additional equipment connected to the medical electrical equipment must comply with the respective IEC
standards (e. g. IEC60950 for data processing equipment, IEC 60601-1 for medical electrical systems). Anybody
connecting additional equipment to medical electrical equipment confi gures a medical system and is therefore
responsible that the system complies with the requirements for medical electrical systems. Attention is drawn
to the fact that local laws take priority over the above mentioned requirements. When more than one device
is connected to use, the accumulation of leakage current may cause the danger of security. If in doubt, consult
your local representative or the technical service department.
5. If the integrity of the external protective conductor in the installation or its arrangement is in doubt, equipment
shall be operated from its internal electrical power source.
Warning!
1. When the device works abnormally, stop working, turn off the power and check the reason, then contact
EICKEMEYER® about it.
2. Turn off power and pull out of the plug from socket after each ultrasonic diagnostic operation.
3. It is forbidden to drag and press the power and probe cables emphatically; regularly inspect whether there is
pull-apart and bareness. If there is a phenomena like this, turn off power supply immediately, stop using it and
replace it with a new one.
4. It is forbidden to load and unload the probe or move the instrument in galvanic to avoid danger of safety.
5. Pull out of the plug from socket after operation in thunderstorm weather to avoid the instrument being
damaged by lightening.
6. If the temperature changes greatly in short time, this will cause vapor recovery inside of instrument, the case
may damage the instrument.
7. The instrument is turned off completely only by disconnecting the power supply from the wall socket.
Warning!
The power adapter, probe, shutter release and power supply cord as described in this section are replaced by the
operator. But these parts must be provided by EICKEMEYER® or his authorized supplier.