40
USER MANUAL
MAGIC 500 PLUS ULTRASOUND UNIT
T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com
11.5 Periodic Safety Checks
To ensure system performance and safety, it must be checked after using 1 year. When check the device, please consult the
International Trade Dept of EICKEMEYER® or its dealers.
Inspect sorts
Inspect items
Cleaning
System inside
Peripheral equipment
Electrical safety
Earth leakage current
Touch current
Patient leakage current
Dielectric strength
Mechanical safety
Check the control panel
Peripheral equipment installation agencies
Other mechanical parts
Probe appearance
Image recording
Images in each mode
Images recorded using a standard probe
12. STORAGE AND TRANSPORTATION
Storage and Transportation
1. If the device is stored over 3 months, take it out of the packing case, connect it to power supply for 4 hours, and then
disconnect the power and place it in the case again following the direction indicated by arrows on the package. Store the
case in the warehouse. Do not pile the case. The device case should have adequate space from ground, walls and ceiling
of the warehouse.
2. Environment requirement
Ambient temperature: -20 °C – 55 °C ; Relative humidity: 30 % – 93 % (without condensation).
Atmospheric pressure: 700 hPa – 1.060 hPa. The warehouse should be well ventilated and free of direct sunlight and
corrosive gas.
3. Shockproof measures have been taken inside the packing case to allow for transport by air, railway, land and sea. The
goods shall not be exposed to poor weather conditions like rain and snow, nor shall the goods be placed upside down,
bumped, knocked or over-stacked.
13. SAFETY CLASSIFICATION
1. Classifi ed according to electric shock protection type:
Class I, internally powered equipment
2. Classifi ed according to electric shock protection degree:
Type B applied part
3. Classifi ed according to the degree of protection against ingress of liquid:
Main unit belong to IPX4 equipment
4. Classifi ed according to operation safety in condition of existence of fl ammable anesthetic mixture with air or oxygen or
nitrous oxide:
It is neither of category AP equipment nor of category APG equipment
5. Classifi ed according to mode of operation:
Continuous operation equipment
6. Classifi ed according to the protection of radio services:
Group I Class A equipment