4x50 EtherNetIP.190129.1v4: User manual
WWW.EILERSEN.COM
Version: 2020-11-24, rev.: 1v4
Page: 15
3.3 Scaling
By use of a DIP-switch it is possible to select the desired scaling of the weight
signals. The scaling of the weight signals on the Ethernet is determined by
SWE.1-2 as follows, where the table shows how a given weight is represented
on the Ethernet depending on switch settings:
Weight
[gram]
SWE.1 = OFF
SWE.2 = OFF
(1 gram)
SWE.1 = ON
SWE.2 = OFF
(1/10 gram)
SWE.1 = OFF
SWE.2 = ON
(1/100 gram)
SWE.1 = ON
SWE.2 = ON
(10 gram)
1,0
1
10
100
0
123,4
123
1234
123400
12
12341
12341
123410
1234100
1234
3.4 Measurement time
By use of DIP-switches it is possible to choose between four different meas-
urement times. All load cells are sampled/averaged over a measurement period
determined by SWE.3 and SWE.4 as follows:
SWE.4
SWE.3
Measurement time
OFF
OFF
20 ms
OFF
ON
100 ms
ON
OFF
200 ms
ON
ON
400 ms
The hereby found load cell signals (possibly filtered) are used on the Ethernet
until new signals are achieved when the next sample period expires.