4x50 EtherNetIP.190129.1v4: User manual
WWW.EILERSEN.COM
Version: 2020-11-24, rev.: 1v4
Page: 32
7.6 Appendix F
– Status Codes
Status codes for the connected load cells are shown as a 4 digit hex number.
If more than one error condition is present the
error codes are OR’ed t
ogether.
CODE
(Hex)
CAUSE
0001
Reserved for future use
0002
Reserved for future use
0004
Reserved for future use
0008
Reserved for future use
0010
Reserved for future use
0020
Reserved for future use
0040
Reserved for future use
0080
No answer from load cell
Bad connection between load cell and 4015 load cell module?
Bad connection between 4015 load cell module and 4040 communication module?
Bad connection between 4040 communication module and 4050 module?
Bad setting of DIP switches on 4040 communication module?
0100
Reserved for future use
0200
Reserved for future use
0400
Reserved for future use
0800
No load cell answer
Bad connection between 4040 communication module and 4050 module?
1000
Normal operation NOT selected
LcSignal(X)
NOT transferred as normal operation is NOT selected!
This is NOT a real load cell error, but an indication that the
Actual Value Selec-
tor
(
AS
) is set to a value different from 0 by the PLC.
2000
Reserved for future use
4000
Reserved for future use
8000
Reserved for future use
Please note that the above listed status codes are valid when the 4040 com-
munication module is equipped with standard program.