EIT Part Number: P/N IM-0110 Rev A Issued April 2019 

MicroCure

®

 Radiometer and 

MicroCure® DataReader  

User Manual 

   

 

 

 

 

 

 

MicroCure

®

 Radiometer and MicroCure

®

 DataReader  

 

 
 
 
 
 
 
 
 
 
 
 

 

 

 

Electronic Instrumentation and Technology (EIT) LLC 

309 Kelly’s Ford Plaza SE 

Leesburg VA 20175 USA  

Phone: 703-478-0700 

Email: 

uv@eit.com

 

Web: www.eit.com 

 

Summary of Contents for MicroCure DataReader

Page 1: ...icroCure Radiometer and MicroCure DataReader User Manual MicroCure Radiometer and MicroCure DataReader Electronic Instrumentation and Technology EIT LLC 309 Kelly s Ford Plaza SE Leesburg VA 20175 USA...

Page 2: ...s 6 Operation 7 Error Messages 8 Maintenance 8 Appendix A MicroCure Product Specifications 11 Appendix B DataReader Product Specifications 12 Appendix C Regulatory Statement 13 Appendix D Technical Su...

Page 3: ...ng applications that would ordinarily be inaccessible to larger radiometers Applications for the MicroCure include Narrow conveyor curing such as credit card processing Small container curing bottles...

Page 4: ...is not exposed to UV above the start threshold point at which the MicroCure starts counting UV within four minutes it will enter sleep mode and not count the current wake up session as a reading o See...

Page 5: ...ry symbol auto shut off and works with all MicroCure units Top Left Data Reader with Display Bottom Left DataReader display showing irradiance value of 1 435 W cm2 in UVA Top Right Top display shows a...

Page 6: ...a UV source is based on the source type UV Broadband Sources are mercury based and characterized by a broad i e wide output across the UV spectrum Typical bulb types include mercury sometimes called...

Page 7: ...s value by measuring the lamps irradiance many times during the data collection run and then summing these values together Energy Density is the total amount of energy exposure over time It is the mat...

Page 8: ...damage both the radiometer and or your equipment Do not allow the two case sides of a MicroCure radiometer to contact each other Doing so will reset the unit and any stored data will be lost Do not a...

Page 9: ...h the optics window side facing the UV source s 6 The MicroCure radiometer must be exposed to UV energy within four minutes after being reset or it will return to a standby sleep mode The UV irradianc...

Page 10: ...rradiance values collected with the MicroCure should be consistent when consistent data collection techniques are followed The irradiance values displayed by the MicroCure are instantaneous peak irrad...

Page 11: ...dip by placing a used swab back into the IPA This prevents prevent contamination of the IPA from any material picked up from the optics o Use the second clean swab in a gentle circular motion to dry t...

Page 12: ...used in inks coatings adhesives and resins that may be near or come in contact with the MicroCure The customer assumes all responsibility for damage to their EIT instrument as a result of chemical co...

Page 13: ...needed to activate the MicroCure Use a MicroCure with the correct operating dynamic range 10 Watt UVA or UVV 50 mW cm2 2 Watt UVA or UVV 10 15 mW cm2 200 milliWatt UVA 1 2 mW cm2 Sample Rate Smooth Mo...

Page 14: ...ivate reset the unit Display 4 digit LCD resolution shown to 0 001 Joule Watt cm24 digit liquid crystal display Dimensions 5 77 X 4 38 x 1 45 139 7 x 108 x 36 8 mm Weight 11 75 oz 333 11 grams Operati...

Page 15: ...326 1 2005 and the specific tests performed were Emissions EN 55011 2007 Radiated Immunity EN 61000 4 3 2006 A1 2008 and Electrostatic Discharge EN 61000 4 2 2009 The MicroCure and MicroCure DataReade...

Page 16: ...der this warranty is limited to replacing or repairing the defective goods at our option EIT shall provide all materials and labor required to adjust repair and or replace the defective goods at no co...

Page 17: ...te packaging if the original packaging is not used The customer is responsible for insuring the unit during transportation to EIT Once received EIT will determine if the issue will be covered under wa...