Further functions
Hemo
Point
®
H2
User Manual
37
- Version -
Software Version
Electronic hardware
Mechanical hardware
- Adaptation
-
Correlation function to conform the output of results with different
devices and/or methods.
y = m * x + n
m … correlation factor
n … offset
if m = 1 and n = 0 (factory defaults) the function
is not active
The correlation can only be enabled by authorized personnel on
special request. Contact EKF Diagnostics’ Technical Service
Department for additional information.
- SPN
-
The current
S
ervice
P
rocess
N
umber, needed by Servicing
5.3.2 Connect
The function of connecting the
Hemo
Point
®
H2
to a PC can be
found under “Menu”, sub-menu “Connect”. The InterLink™ Lite
software program and a PC cable are required for this function to
operate properly (see Section 10.1 “Replacement parts and
consumer materials” for ordering information). Please refer to the
InterLink™ Lite Software User Manual for further instructions.
5.3.3 Setting the date
The function of setting the current date can be found under
“Menu”, sub-menu “Date”. Setting the date is important for the
correct relationships of the results from the memory (data
storage) or for printing out the results.
The format to enter the date depends on the date format setting
(see Section 5.3.7.2 “Setting date format”).
The arrow buttons increase or decrease
the numerical value in each case.