Important information
Hemo
Point
®
H2
User Manual
6
Follow the user manual!
Each time the analyzer is used, precise knowledge and attention
to these operating instructions is required. Only use the
Hemo
Point
®
H2
system for the purpose which will be described in Section 2.
Danger of fatal electric shock!
•
Under no circumstances should you open the AC power
adaptor. There are no elements inside which require
servicing or maintenance.
•
Never use a mechanically damaged AC power adaptor –
live connections might be exposed.
•
The AC power adaptor is not waterproof. Therefore, never
let liquids come into contact with it. A
lightly
dampened
cloth, however, can be used to clean it when
disconnected. Please see the notes on care in Section 6.
•
Only use the AC power adaptor in a socket that has been
properly installed.
Check that the AC voltage and frequency printed on the AC power
adaptor label match your electrical socket and whether the shape
and configuration of the plug contacts are compatible for this
connection.
Not to be used in areas where there is a risk of explosion!
The instrument is not approved for use in areas where there is a
risk of explosion.
Keep the analyzer away from liquids!
The analyzer is not waterproof. Fluids entering the instrument
could destroy the electrical and optical components in the
analyzer. A
lightly
dampened cloth, however, can be used for
cleaning. Please see the notes on care in Section 6.
Allow the instrument to reach room temperature!
Particularly on changing from a cold into a warm environment (i.e.
after storage or transport) condensation can form inside and on
the outside of the instrument. Wait an appropriate time (approx. 1
hour), before you connect to the main power supply or switch on
the instrument.
WARNING!
DANGER!
WARNING!
DANGER!
WARNING!
WARNING!