Section 4 -
Page 4 of 8
3121-9001-0167_EN.03.00_QT_Manual
04/2016
Step 8
Pull the red slide towards the front of
the analyzer and close the door. The test
will start automatically.
You will be prompted to scan the
Operator and Patient ID, if these
features have been activated.
Note:
If the analyzer displays “Scan
Lot ****** Data Card” then scan the
calibration barcode (see Step 2).
Step 9
When the test has finished the result will
be displayed on the screen and printed
on the (optional) printer, if connected.
Open the door and push the red slide
towards the rear of the analyzer. Remove
the used cartridge and dispose of it as
clinical waste.
Upon closing the analyzer door, the
display will return to the home screen
and the analyzer will be ready to run
another sample.
Note:
See Section 4 -
Retrieving a
Result from the Analyzer Memory
9
8