Section 4 -
Page 6 of 8
3121-9001-0167_EN.03.00_QT_Manual
04/2016
7.1% A1C
DCCT
54 mmol/mol IFCC
The Test Result
When the test is finished, the result will be displayed on the screen in the specified
units (Section 3 -
Reported Units
). The image below shows an example of the result
of a Quo-Test A1C Test. In this case, the test result was 7.1 % A1C. The letters “DCCT”
below the test result show the selected primary reported units of the test at the time
it was run.
Important
: make a note of the calibration scheme in order to accurately
compare results. Below the main result is a derived result showing the equivalent value
in mmol/mol IFCC units.
The result will stay on the screen until
the analyzer door is opened.
The result
is stored in the analyzer memory (see
Section 4 -
Retrieving a Result from the
Analyzer Memory
). On removal of the
used cartridge, the analyzer will be ready
to perform another test.
Printing the Test Result
If you have the optional printer
connected, the result will be printed
when the test has finished. The result is
printed onto a self-adhesive label to be
kept with the patients’ records.
Time:
Date:
Result:
Lot:
Inst ID:
Test ID:
Patient:
Operator:
Quo-Test A1C
11:34
21/06/13
7.1 % A1C DCCT
54 mmol/mol IFCC
020130
021132
00053
xxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxx