Section 6 -
Page 2 of 4
3121-9001-0167_EN.03.00_QT_Manual
04/2016
MESSAGE OR
PROBLEM
DESCRIPTION
RESOLUTION
Error 04/05
Part of the optical system has
failed
Remove the old cartridge, if present, close
the door and allow the analyzer to return
to the home screen*
Error 06 / 07
The x- or z-axis motor has failed
Power off the analyzer, wait 10 seconds,
and power on again
Remove the old cartridge, if present, close
the door and allow the analyzer to return
to the home screen*
Error 08
The stirrer motor has failed
Error 09
The internal fan
has failed
Error 10/11
The x- or z-axis motor has
jammed
Error 12
The stirrer motor has jammed
Error 98
The measurement system
has failed
Contact your local distributor or customer
support
Error 99
The analyzer memory is full
Door Opened
Assay Abandoned
(Error 101)
The analyzer door was opened
before the test had finished
Do not open the analyzer door whilst
a test is running*
Reagent Out of Range
(Error 102 / 103)
The test cartridge is faulty
Remove cartridge and repeat test
with a new cartridge*
Too Little Blood
(Error 104)
Either the blood sample had
a low level of hemoglobin, the
blood collector was under filled
or the filled blood collector was
pushed into the cartridge prior to
running the test
Remove cartridge and repeat test
with a new cartridge, making sure that the
blood collector is filled correctly and that it
is carefully placed into the cartridge*
(see Section 4 -
Running a Test
)
Too Much Blood
(Error 104)
Either the blood sample had
a high level of hemoglobin, or
blood collector was overfilled
Blood Mix Failure
(Error 105)
The blood sample did not form
a homogenous solution with the
buffer
Remove cartridge and repeat test
with a new cartridge*
Invalid Result
(2)/(1)
(Error 106)
A result was not able to be
calculated
Remove the cartridge and repeat
with a new cartridge*
*
If the problem continues, contact your local distributor or customer support.